40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
3A/m
Compliance level
±6 kV contact
±8 kV air
±2kV for
power supply lines
±1 kV
differential mode
<5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT)
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
3A/m
Electromagnetic
environment – guidance
Floors should be wood,
concrete or ceramic tile. If
floor are covered with
synthetic material, the
relative humidity should be
at least 30%.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the TMB-1018 requires
continued operation during
power mains interruptions,
it is recommended that the
TMB-1018 be powered
from an uninterruptible
power supply or a battery.
Power frequency magnetic