CAUTION

. Check chemical resistance of the wetted parts against the used chemicals 

before using the device. 

 

 

CAUTION

. The flow measuring cell is suitable for a maximum operating pressure of 

10 bar. Make sure that the tubing, you want to use, is also suitable for the expected 
pressure. 

 

 

CAUTION

. Depending on the configuration of the internal or an external light source 

optical radiation may leak, that can cause eye and skin damage. Note the safety 
information concerning wavelength and optical power. When using an external light 
source read the associated manual of the manufacturer and follow the safety instructions 
therein. 

6.1

 

(Dis)Assembly of the measuring cell 

Both  Qmix  λ  modules  have  a  holder  for  the  measuring  cell.  The  only  functional  difference  is  the 
presence of spring contacts on the small which serve to power the LED light source. 

The measuring cell consists of the same parts on both Qmix λ modules - two connection pieces and an 
intermediate piece. The assembly of the measuring cell on the devices will be described below. 

Light source 

Window 

Window 

Fluid 

Spectrometer 

Summary of Contents for Qmix L

Page 1: ...QMIX Hardware Manual ORIGINAL MANUAL 3 00 FEBRUARY 2016...

Page 2: ...2 Qmix Hardware Manual CETONI GmbH Wiesenring 6 07554 Korbussen Germany T 49 0 36602 338 0 F 49 0 36602 338 11 E info cetoni de www cetoni de...

Page 3: ...rm the purchaser of any alterations new developments or improvements to the software or make these available to him No legally binding assurance of certain features is given CETONI GmbH is not liable...

Page 4: ...4 Qmix Hardware Manual...

Page 5: ...al data 8 3 3 Electrical data 8 3 4 Interfaces 8 3 5 Wetted parts 9 4 Use 10 4 1 General description of the device 10 4 2 Intended use 10 4 3 Reasonably foreseeable faulty application 10 4 4 Safety ad...

Page 6: ...inor changes 1 10 12 09 2013 Added Qmix BaseXT and TC power consumption Q 1 11 21 08 2014 Adaptation of the maximum heating temperatue of the Reaction module Q heating column and the High temperature...

Page 7: ...rdware Manual 7 2 Scope of supply The following items should be included in the scope of supply QMIX MODULE Spectrometer as ordered Light source as ordered OCEANVIEW SPECTROSCOPY APPLICATION MANUAL HA...

Page 8: ...ondensing STORAGE HUMIDITY 20 90 non condensing 3 2 General data DIMENSIONS L X W X H 310 x 75 x 90 100 mm WEIGHT 0 9 kg 1 1 kg WAVELENGTH RANGE Depending on the configuration MAXIMUM PRESSURE 10 bar...

Page 9: ...Qmix Hardware Manual 9 3 5 Wetted parts WINDOW fused silica SiO2 LIGHT SOURCE PPS polyphenylene sulfide MEASUREMENT SECTION PTFE tubing SEALS FPM...

Page 10: ...e omitted CAUTION The unit must not be used as a medical device or for medical purposes 4 4 Safety advice For the safe operation of the Qmix modules it is necessary to observe the safety measures from...

Page 11: ...original packaging Use the original packaging for transportation or shipping of the module Concerning the storage conditions please observe the data from chapter Technical data CAUTION Risk of damagi...

Page 12: ...ge of 350 to 1100nm It provides an integrated LED light source which can be adapted to the required wavelength range An adapter with SMA 905 connector is also available The larger module can be adapte...

Page 13: ...Note the safety information concerning wavelength and optical power When using an external light source read the associated manual of the manufacturer and follow the safety instructions therein 6 1 D...

Page 14: ...window has to be placed with this side first Attach the connection piece to the holder on the module with two M2 5x8 screws using a 2 mm Allen key It may be helpful to hold the module upright lest the...

Page 15: ...sides The PTFE tubing can be used optionally Finally the second connection piece is mounted When the shortest intermediate piece is used the final connection piece is screw mounted directly on the exi...

Page 16: ...t is your decision which of the ports you would like to be in or outlet 6 3 Light source The small module provides an integrated LED light source The brightness of the LED can be adjusted to your need...

Page 17: ...wo marked screws with a phillips screwdriver and then simply exchange the LED attachment Please do not remove the smaller screw in the centre If you want to mount the SMA 905 adapter instead of the LE...

Page 18: ...18 Qmix Hardware Manual Mounting plate SMA 905 adapter Window and O Ring...

Page 19: ...mbH to coordinate further actions The device may only be opened by CETONI GmbH or thereby authorized service staff Otherwise the warranty and guarantee claims are void Software related troubles are de...

Reviews: