- 6 -
7.
Removing the sensor
The sensor is removed in the following steps:
•
Disconnect the sensor cable from the BB2 and the flushing hose from solenoid.
•
Open the clamp, and pull the rod out of the mounting bracket
•
Make sure all water inside the rod is drained.
•
Clean the sensor and rod with a brush or clean cloth. Do not use a wire brush!
•
Open the black sensor adapter.
•
Flush the inside of the rod with clean water.
•
Mount the protective cap (or a small plastic bag) on the sensor cable connector.
•
Pull the cable and flushing hose out of the rod.
•
Blow compressed air through the flushing hose to get rid of the water in the hose
and the sensor.