Cerlic ITX 20 Manual Download Page 5

                                                                                                        

 

 

- 5 - 

 

6.

 

Mounting of ITX Suspended Solids Sensor 

The ITX is mounted on a telescopic fiberglass rod and placed in a mounting bracket that 
fastens to a handrail. Please refer to Appendix 1. As an alternative way of mounting the 
sensor, an adjustable slide rail holder is available, please refer to the Appendix 2 for 
further details. 

When the sensor is measuring in a fume, it is important to find a place where the 
suspended solids concentration is representative. 

Make sure the flushing nozzle is downstream from the lenses pointing against the 
stream. This will avoid having the measurement disturbed by turbulence from the 
nozzle. At the same time it will produce a shield around the nozzle, due to a constant 
over pressure, preventing particles from getting in.  

Cable Connections  

Connect the sensor to the BB2 control box using the attached connector on the end of 
the sensor cable. In the event that two sensors are to be connected to the same BB2 
control box, use the Y-splitter. 

NOTE! 

Never try to turn the connector housing to fasten or remove the connector, only 

the fluted ring at the top of the connector shall be turned.  

Installation Tips 

 

 

Adjust the rod so that the sensor is at least 12” below the liquid surface or the lowest 
water level in decant applications to prevent the sensor from coming out of the 
liquid.   

 

In an aeration tank, ensure that the sensor is not directly above a diffuser head.  It 
should be on the backside of the rolling diffuser effect.   

 

Flushing may not be required if the tank is well agitated.  To verify the need for 
flushing, remove the sensor from the liquid after it has been in the liquid for several 
days.  Place the sensor in a bucket of clean water and note the concentration value.  
If the value does not get close to zero, then the flush system should be utilized.   

 

When installing in a clarifier, compressed air flushing is required due to no agitation 
of liquid and to remove oil and grease film on lens.  This is especially applicable in 
primary clarifiers.   

 

When using the ITX for influent applications, always install the unit after the bar 
screen.  If the bar screen spacing is larger than ¼ ’ (6 mm), then a baffle or diffuser 
plate should be installed in front of sensor to prevent rags from catching on the 
sensor head.  On influent applications, compressed air flushing is recommended due 
to the oil or grease in the liquid.   

 

In effluent applications a CTX type of sensor is recommended, it has better 
resolution at low concentrations. 

Summary of Contents for ITX 20

Page 1: ...2002 11 19 MANE11 02 Ver 2 00 ITX 20 Suspended Solids Sensor...

Page 2: ...g plate for cleaning solenoid valves 8 9 Menu for ITX sensor 9 Settings 9 Calibrate 9 System 10 10 Calibration 11 Getting good measurement 11 Calibration screen 11 Zero Calibration 12 Entering sample...

Page 3: ...ntenance under critical applications The measuring lenses within the housing are made of glass The electronics and optics are protected in the rugged casing ensuring its reliability in very demanding...

Page 4: ...check that the content corresponds to your order and packing list Every shipment should include ITX Sensor w 10 m 33ft cable and flushing hose P N 11305455 Optional parts for ITX Suspended Solids Sens...

Page 5: ...lation Tips Adjust the rod so that the sensor is at least 12 below the liquid surface or the lowest water level in decant applications to prevent the sensor from coming out of the liquid In an aeratio...

Page 6: ...re all water inside the rod is drained Clean the sensor and rod with a brush or clean cloth Do not use a wire brush Open the black sensor adapter Flush the inside of the rod with clean water Mount the...

Page 7: ...e the control box For example if the solenoid is wired to relay 1 select 1 for flushing Next time displays the next time flush will be activated pushing ENTER will set it to current time and those sta...

Page 8: ...in flushing nozzles Cerlic offers a predrilled Aluminum handrail mounting plate The Mounting plate is fixed to the rail using two u bolts Please see technical specifications for details and ordering i...

Page 9: ...ing cycles Length sec 0 999 seconds duration of flushing cycle Freeze sec 0 999 seconds extra freeze time of output signal after a flushing cycle Relay 1 or 2 Select relay to operate solenoid for flus...

Page 10: ...e SA 1 SA value for sample 1 Cons 1 Lab test consistency or suspended solids value for Sample 1 SA and Cons repeated for samples 2 to 5 Info Press ENTER to go to info read only menu This menu is for C...

Page 11: ...es while new calibrations will start from scratch Calibration screen The ITX sensor information menu is the calibration curve screen To change between the main menu and the calibration screen press an...

Page 12: ...ect unit consistency suspended solids is selected for the application Select unit in the sensor menu SETTINGS UNIT Remove the sensor from the process clean the sensor head and dip it in a bucket of cl...

Page 13: ...et so the sensor is at least 12 36 cm below the lowest liquid level in the basin Take the bucket to the lab for analysis Note the concentration of the sample determined at the lab Enter sample 1 conce...

Page 14: ...to set high and low alarms values that can be used to notify when solids have reached critical points MAX sets the 20 mA point output MIN sets the 4 mA point output may be a negative value for specia...

Page 15: ...loped and manufactured in accordance with the following standards SS EN50 082 2 immunity SS EN50 081 2 emission SS EN61 010 1 safety 89 336 EEC 92 31 EEC 93 36 EEC 73 23 EEC Optional parts for ITX Sus...

Page 16: ...16 13 Dimensions 2 60 8 94 10 04 2 60 66...

Page 17: ...cket Make sure that the bracket guide tracks are properly seated in the bracket Fasten the safety locking clamp Check that the bracket is safely fixed to the rail for the spring to work the way it is...

Page 18: ...18 Appendix 2 Assembly of adjustable slide rail...

Page 19: ...me Use or to select the sensor in the main display Go to the sensor menu it is accessed by pressing ENTER for five seconds Select System and press ENTER Type __________________________________________...

Page 20: ...___________________________________________ Cons 1 _____________________________________________________ Ch1a _____________________________________________________ Ch1 ________________________________...

Page 21: ...________________________ Alarm Relay _____________________________________________________ Analog _____________________________________________________ Cleaner ________________________________________...

Reviews: