Cerlic 11305473 Manual Download Page 17

 

 

 

                 

 

- 17 - 

 

 

Appendix 1, Assembly of handrail mounting kit 

 

 

 

 

 

Disassemble the bracket guide and place 
around the telescopic rod.  

 

Use the SS screws on bracket to tighten the 
bracket to the rod. 

 

Mount the Flexible Mounting Bracket on existing handrail or on 
separate holder. The bent lip on the mounting plate shall be on top 
and faced toward the liquid or tank. Adjust the Mounting Bracket to 
the correct angle and tighten the nuts. The bracket shall be fixed to 
the rail, and must not be able to rotate around it. 

 

Insert the PVC bracket guide with the telescopic rod 
into the mounting bracket. Make sure that the bracket 
guide tracks are properly seated in the bracket. 

 

Fasten the safety-locking clamp. 

 

Check that the bracket is safely fixed to the rail for the 
spring to work the way it is intended. 

Example of a holder 
to be used for the 
mounting bracket 
when no handrail is 

 

Pull the cable and hose through the sensor holder and rod. 

 

Connect the telescoping fiberglass rod to the sensor with the two piece black PVC adapter.  

 

The adapter halves should be tightened until snug, which will leave about 1/16" gap.  

NOTE!

 The gap is required so the water can drain from the rod.  

 

Adjust the length of the telescopic sensor rod as necessary by twisting the nuts while holding the 
rod. 

NOTE!

 Do not extend the rod sections beyond the black lines. This could lead to rod damage. 

Summary of Contents for 11305473

Page 1: ...2002 11 18 MANE11 02 Ver 2 00 O2X Dissolved Oxygen Sensor...

Page 2: ...Installing a new electrode 7 Mounting the protection plate 7 9 Cleaning unit 8 Cleaning the Flushing Nozzle 8 Mounting plate for solenoid valves 9 10 Menu for O2X Transmitter 10 11 Calibration 12 Zer...

Page 3: ...t in flushing nozzles allow for the most accurate readings with little maintenance The electronics is protected in the rugged casing ensuring its reliability in very demanding environments The sensor...

Page 4: ...hould include O2X sensor P N 11305473 including 33 ft 10 m cable and 33 ft 10m flushing hose You will also need an electrode below to get the sensor working Optional parts for O2X Dissolved Oxygen sen...

Page 5: ...liquid surface or the lowest water level in decant applications SBRs to prevent the sensor from coming out of the liquid In an aeration tank ensure that the sensor is not directly above a diffuser he...

Page 6: ...ing a new electrode The electrode can usually be both assembled and disassembled without tools Do not ever clamp onto the sensor housing or part of the cable when installing or removing the flushing h...

Page 7: ...nting Tighten the flushing head and mount the protection plate if needed See below for instructions on mounting the protection plate A new air calibration must be performed whenever the electrode is c...

Page 8: ...olenoid is wired to relay 1 select 1 for flushing Next time displays the next time flush will be activated pushing ENTER will set it to current time and those start cleaning Cleaning the Flushing Nozz...

Page 9: ...lushing nozzles Cerlic offers a predrilled Aluminum handrail mounting plate The Mounting plate is fixed to the rail using two u bolts Please see technical specifications for details and ordering infor...

Page 10: ...to be seen as additional information not a primary measurement Cleaning Press ENTER to go to Cleaning program Cleaner None Flush or Brush Do not select Brush since this does not exist for O2X sensor I...

Page 11: ...ly SoftW Software version of the sensor read only Temp Sensor temperature read only Info Press ENTER to go to info read only menu This menu is for Cerlic internal use it may change without notice MS0...

Page 12: ...brating a sensor Air calibrating a sensor The most important thing to keep in mind while performing an air calibration is that it is necessary to create a constant environment The electrode cartridge...

Page 13: ...ar hPa 29 2 990 29 3 993 29 4 997 29 5 1000 29 6 1003 29 7 1007 29 8 1010 29 9 1014 30 0 1017 30 1 1020 30 2 1024 30 3 1027 30 4 1030 The following appear in the Calibration menu as read only Temp inf...

Page 14: ...to set high and low alarms values that can be used to notify when the level has reached critical points MAX sets the 20 mA point output MIN sets the 4 mA point output may be a negative value for speci...

Page 15: ...SS EN50 082 2 immunity SS EN50 081 2 emission SS EN61 010 1 safety 89 336 EEC 92 31 EEC 93 36 EEC 73 23 EEC Optional parts for O2X Dissolved Oxygen Sensor Electrode cartridge P N 11805160 Mounting br...

Page 16: ...16 14 Dimensions...

Page 17: ...cket Make sure that the bracket guide tracks are properly seated in the bracket Fasten the safety locking clamp Check that the bracket is safely fixed to the rail for the spring to work the way it is...

Page 18: ...18 Appendix 2 Assembly of adjustable slide rail...

Page 19: ...____________ Box heat _____________________________________________________ Leave the BB2 menu by pressing and ENTER at the same time Use or to select the sensor in the main display Go to the sensor m...

Page 20: ...Relay _____________________________________________________ Analog _____________________________________________________ Cleaner _____________________________________________________ Cleaning interva...

Reviews: