Step 9)
Check the Column Collar for proper adjustment. The Column Collar should sit
evenly on the Column (#1C).
Step 10)
Locate the Table Clamp (part #13C) and insert it into the Table Support (#7C). Once inserted,
tighten it by hand. Once this is done, rotate the Worm Elevation (#18C) clockwise with the
Crank (part #8C). If necessary, loosen the Screw (#11C) in the Crank and reposition the
Crank as close to the Table Support as possible. Tighten the Screw (#11C).
Head to Column
Step 1)
Insert the Locking Shoe (part #15B) into place inside the bottom of the Guide Column
(#18B) - see Figure 2.
Step 2)
Insert the second Locking Shoe (Part #15B) into the Guide Column (part #18B)
on the inside of the hole designated for the upper Clamping Lever (part #30B).
Step 3)
With assistance, slide the Guide Column (part #18B) onto the Column (#1C) as
shown in Figure 2. It may be necessary to loosen the Guide Pin (part #16B) as well as the
Clamping Levers (part #30B) slightly to ensure that they do not protrude internally, as this will
prevent the Guide Column from sliding fully into position. Align the Guide Column (#18B) with
the Base (#6C) and firmly secure it into position with the Clamping Levers (#30B).
Step 2)
If it is not in place, screw the Depth-Lock Screw (part #38B) into the Depth Ring
Stop (part #37B) - see Figure 3.
Step 3)
Screw the Hub (part #36B) into the Ring-Depth Stop (#37B). Screw all three (3)
Feed Rods (part #35B) firmly into the Hub (#36B). Screw each of the three Feed
Knobs (part #34B) onto the Feed Rods (#35B) - see Figure 3.
Table Support (7C)
Guide Column (18B)
Clamping Lever (30B)
Locking Shoe (15B)
Figure 2 - Head to Column
Depth Lock Screw (38B)
Hub (36B)
Feed Rod (35B)
Depth Ring Stop (37B)
Feed Knob (34B)
Figure 3 - Feed Rod Assembly
#42289
Page 6
Brought to You by Augusta Flint