AUTORANGING INSULATION TESTER

                     

 

instruction manual

 

 

I.            SAFETY INFORMATION 

 

Read the following safety information carefully before attempting 
to operate or service the meter. 

 

The circuit under test must be de-energized and isolated before 
connections are made except for voltage measurement. 

 

Circuit connections must not be touched during a test. 

 

After insulation test, capacitate circuits must be allowed to 
discharge before disconnecting the test leads. 

 

To avoid damages to the instrument do not apply the signals, 
which exceed the maximum limits shown in the technical 
specification tables. 

 

Do not use the meter or test leads if they look damaged. Use 
extreme caution when working around bare conductors or bus 
bars. 

 

Use the meter only as specified in this manual; otherwise, the 
protection provided by the meter may be impaired. 

 

Caution when working with voltages above 60V DC or 30 V AC 
RMS. Such voltages pose a shock hazard. 

 

Before taking resistance measurements or testing acoustic 
continuity, disconnect circuit from main power supply and all loads 
from the circuit. 

Environment conditions: 

 Installation Categories III 600 

V

 

 Pollution Degree 2 
 Altitude up to 2000 meters 
 Indoors use only 
 Relatively humidity 80% max. 
 Operation Temperature 0

40

 

Maintenance & Cleaning: 

 Only qualified personnel should perform repairs or servicing not 

covered in this manual. 

 Periodically wipe the case with a dry cloth. Do not use 

abrasives or solvents on these instruments. 

Summary of Contents for 360

Page 1: ...OFF 600V 1000V I CAT CAT II I I Lo V 1000V M Hi 4000 1000V 4000M V TEST 4000M 250V 500V 4000M M 0 1000 V 2000 3000 360 ...

Page 2: ...ERAL SPECIFICATION 2 III ELECTICAL SPECIFICATION 3 IV SYMBOL DEFINITION BUTTON LOCATION 4 V MEASURING FUNCTION 5 1 ACV Function 5 2 DCV Function 5 3 Low Ohm Function and Continuity Function 5 4 MegaOhm Function 5 VI AUTO POWER OFF 6 VII BATTERY CHANGING 7 VIII FUSE CHANGING 7 AUTORANGING INSULATION TESTER DEC 1998 ...

Page 3: ... if they look damaged Use extreme caution when working around bare conductors or bus bars Use the meter only as specified in this manual otherwise the protection provided by the meter may be impaired Caution when working with voltages above 60V DC or 30 V AC RMS Such voltages pose a shock hazard Before taking resistance measurements or testing acoustic continuity disconnect circuit from main power...

Page 4: ...segments analog Bar Indication Over range Indication OL will be shown on the LCD Panel when out of range measurement is made Low Battery Indication The will be show when the battery need to be changed Sampling Rate 2 5 times sec Digital Display 10 times sec Bar Graph Power Source 1 5V AA size Battery 8 Operating Temperature and Humidity 0 to 40 32 to 104 bellow 80 RH Storage Temperature 10 to 60 1...

Page 5: ...lution Accuracy 400Ω 0 1Ω 4000Ω 1Ω 1 rdg 5dgts Continuity beeper Range Active Protection 40Ω 600Vrms MΩ Autoranging Range Resolution Accuracy 4 40 400 4000MΩ 250V 4 40 400 4000MΩ 500V 4 40 400 4000MΩ 1000V 1KΩ 3 rdg 5dgts 2GΩ 5 rdg 5dgts 4GΩ Power Consumption Battery 9 5 V Range Condition mA approx ACV 16mA DCV 16mA Ω 22mA Ω 0Ω 190mA 250V 50mA 250V 250K 120mA 500V 60mA 500V 500K 150mA 1000V 85mA 1...

Page 6: ...AL SHOCK OPENING CASE SIZE AA 1 5VX8 REMOVE TEST LEADS BEFORE F 0 5A 600V M 1 2 3 6 5 4 8 7 42000 63000 84000 21000 0 IV Symbol Definition and Button Location LCD Display TEST Button Function Selector Power Lock Button Data Hold Button Hi Input Terminal Lo Input Terminal Battery Compartment Cover Low Battery High Voltage Output Lock Enable Data Hold Continuity Beeper Enable Analog Bar Scale Analog...

Page 7: ...o the Hi terminal Connect test lead to the test circuit in parallel If the reading is less than 40 Ohm the continuity beeper will sound One may null the lead resistance under 40 Ohm by shorting the test lead than press the test key When the lead resistance is recorded a ZERO symbol will display on the LCD Press TEST key again to go back to normal operation If the lead resistance is great than 40Ω ...

Page 8: ...ltage output Press the test button again to stop the test voltage output A series of beeping with shorter period indicate the discharging in progress When the beeper stops the discharge is completed The test result will be held on the display automatically If test period extend longer that 3 minute the test source will shot down automatically Caution Do not activate the test before the lead is con...

Page 9: ...with a screwdriver Eight AA 1 5V batteries are needed to replace the old ones After all Batteries are changed put the cover back and fasten the screw VIII FUSE CHANGING When connect the meter under Ohm rang to a source 10V the protection fuse will break the circuit and new one will be needed for replacement To replace the fuse user should prepare an F 0 5A 600V fuse First the user should turn the ...

Page 10: ...CENTER TECHNOLOGY CORP 4 F NO 415 Jung Jeng Rd 238 Shu Lin Chien Taipei Taiwan TEL 886 2 26763926 E Mail center centertek com FAX 886 2 26763925 http www centertek com 360 01 ...