CAUTION 

- Indicates a potentially hazardous situation which, if not avoided, may result in minor 

or moderate injury. It may also be used to alert against unsafe practices. This signal word is to be 
limited to the least dangerous situations. 
 

1.2.1  Laser Safety

 

 

 

 

WARNING 

There is a risk of an incorrect diagnosis or treatment.

 The isocenter of the lasers 

must coincide with the isocenter of the diagnostic or treatment equipment. Testing the alignment of  the 
lasers to the isocenter of the diagnostic or treatment equipment is based on the hospitals procedures. Newly 
installed equipment can possibly shift as building walls and floors settle and as building materials dry out. 
If the equipment or the lasers have been moved, it is advised to realign the  lasers to the isocenter before 
use. 

 

 

CAUTION 

The laser light is a risk for eye damage.

 The laser used in the MAXX-700 patient 

positioning system is a Class 2 laser. To prevent eye damage, do not allow patients to stare into the laser 
beam. The normal blink reflex will prevent eye damage. If a patient is unable to blink normally, then the 
laser operator must prevent the laser from shining into the patient's eye. 

 

1.2.2  Safety Labels 

 

 

 

Three safety labels are placed on each MAXX-700 laser unit in the locations shown below: 

 

 

 

 

1. At the upper left corner 

on the left side of the 

cover. 

2. Right under the 

window on the front side 

of the cover. 

3. At the upper right 

corner on the right side of 

the cover. 

Summary of Contents for MAXX-700

Page 1: ...The MAXX 700 User Installation Manual Version 2 0 January 2020...

Page 2: ...e installation 8 2 2 Power Supply 8 2 3 Typical Room Layout Example 10 2 4 Location of Installation 11 2 5 Powering Up The MAXX 700 System 12 3 Troubleshooting 13 3 1 Operation Failure 13 3 2 Other Pr...

Page 3: ...e control to operate the unit 1 1 Intended Use The MAXX 700 lasers are intended to be used to align patients for medical diagnostic imaging and radiation therapy They are intended to be used in a clin...

Page 4: ...ls and floors settle and as building materials dry out If the equipment or the lasers have been moved it is advised to realign the lasers to the isocenter before use CAUTION The laser light is a risk...

Page 5: ...ending on whether the lasers are being used in Europe or North America Safety Labels For Europe The wavelength is 450 nm for a blue laser The wavelength is 532 nm for a green laser The wavelength is 6...

Page 6: ...de The PAN ID The PAN ID is the network identifier number of the wireless network within a room The remote control and all of the laser units in the room must have the same PAN ID to communicate with...

Page 7: ...ir Beam Fan Out Crosshair Tolerance Line Straightness 60 in each axis 90 14 4 arc seconds 0 05mm deviation over 3m at 3m distance Electrical Power Supply Input 100 240V AC 50 60Hz 0 8A Output 12V DC 1...

Page 8: ...the MAXX 700 S Sagittal are used with different systems A typical system consists of Two MAXX 700 CH crosshair lasers mounted on each side beam crosshairs along the X axis One MAXX 700 CH crosshair l...

Page 9: ...tion Some facilities prefer to connect all the lasers to one wall switch so all of the lasers are able to be turned on and off together from a convenient location Some facilities connect the lasers to...

Page 10: ...Page 10 2 3 Typical Room Layout Example Unit 1 Unit 2 Unit 3 Unit 4...

Page 11: ...cket for the front can be used to install all of the units This is highly recommended for ceiling mounted crosshair lasers MAXX 700 CH as well as sagittal lasers MAXX 700 S Figure 2 6 Adjustable Mount...

Page 12: ...re 2 7 Powering Up A MAXX 700 A The Power LED on the bottom left side should be on immediately B The Unit Number LED displays the unit number C The laser starts beaming and the brightness of the laser...

Page 13: ...this is not the case go to the next step 4 Shut down all of the units and power down the remote control Wait at least 10 seconds then power up the laser units followed by the remote control The netwo...

Page 14: ...e switched off and the power must be disconnected from the supply using the appliance inlet WARNING No modifications may be made to the remote control as this will result in the Cemar Electro Inc warr...

Page 15: ...ste upon disposal For proper treatment recovery and recycling please contact your disposal local supplier for further information In a case that a laser fails to function properly please contact our c...

Page 16: ...igations or rights including without limitation any other warranties expressed or implied e g implied warranties of merchantability or fitness for a particular purpose Under no circumstances shall Cem...

Page 17: ...Cemar Electro Inc service representatives are available to work with facility planners architects and hospital personnel throughout the planning and installation process Service department hours are...

Page 18: ...Type Compliance Declaration The MAXX 700 series which comprises wireless communication equipment complies with Directive 2014 53 EU The full text of the EU Declaration of Conformity is available at th...

Page 19: ...treatment recovery and recycling please contact your dealer or supplier for further information Date of Manufacture Manufacturer Part Number Serial Number Lot Number Authorized Representative Europe D...

Reviews: