17
System Installation
8.4.
Scroll to the “Power” section and select “Line Voltage.”
• 60 Hz - select 208 V or 230 V
• 50 Hz - select 200 V, 220 V, or 240 V
8.5.
Select “Ok” and the Save icon in the upper right corner of the screen.
9.
Verify iWave sensor temperature.
9.1.
Select the System Settings icon in the bottom right corner of the screen.
9.2.
Select Tools.
9.3.
Select the “System” tab. The iWave Temperature(s) will appear under “Status”.
NOTE
The “Side iWave Temperature (ºC)” will only appear for the iPrep 16 vessel set.
9.4.
Ensure sensors are clean and free of debris.
9.5.
Verify the temperature is reading correct temperature at room temperature.
9.6.
Press the Home icon in the lower left corner of the screen to return to the Home page.
10.
Calibrate the turntable position.
10.1.
Install the MARSXpress turntable. If a MARSXpress turntable is not available, use the turntable calibra-
tion tool (P/N 404390).