5-2 DC/AC Millivolt Measurements
CAUTION:
5-3 AC/DC Current Measurements
CAUTION:
Do not measure DC/AC voltages if a motor on
the circuit is being switched ON or OFF.Large voltage surges
may occur that can damage the meter.
Rotate the function switch to the
mV DC-AC position.
Insert the black test lead banana
plug into the negative COM jack.
Insert the red test lead banana plug
into the positive V jack.
Press the MODE button to select
"DC" or "AC" millivolts.
Touch the black test probe tip to
the negative side of the circuit.
Touch the red test probe tip to the
positive side of the circuit.
Read the voltage in the display.
Do not make 20A current measurements for longer
than 30 seconds. Exceeding 30 seconds may cause damage
to the meter and/or the test leads.
Insert black test lead banana plug
into the negative COM jack.
For current measurements up to
5000μA DC,set the function switch
to the μA position and insert the
red test lead banana plug into the
μA/mA jack.
For current measurements up to
500 mA DC,set the function switch
to the mA position and insert the
red test lead banana plug into the
μA/mA jack.
•
•
•
•
•
•
•
•
True RMS Autoranging Digital Multimeter
OFF
Auto
Power
Off
10A
mA
μA
Ω
CAP
Hz
%
Temp
°C
°F
V
mV
RANGE
REL
PEAK
MODE
MAX
MIN
HOLD
COM
10A
mA
μA
CAT III 1000V
CAT IV 600V
20A for 30 sec MAX
every 15 min
V
CAP
Ω
Hz%
Temp
10A
500mA
FUSED