PREPARING THE ULTRAMIST

 SYSTEM

Allow the saline to run through the applicator tubing until 

saline is noted at the distal tip.  Close the tubing clamp. 

10

If utilizing the adjustable hook, position generator near the 

edge of a flat work surface (i.e. tabletop or cart).  Extend 

the hook beyond the edge and attach the saline bag.  

Adjust position as necessary so the saline bag hangs freely 

and the generator is stable.

 

11

The display screen will ask for the saline pump to be 

loaded.  Open the saline pump door.  

12

Loop the applicator tubing from the saline bag through 

the saline pump from left to right, center on tubing 

guides, and close the pump door.   Leave enough tubing 

to allow the saline bag to hang freely with the majority of 

the tubing between the pump and treatment wand.

13

Open the  tubing clamp.  Saline will not flow until the 

treatment on/off button is initiated. 

14

The system is now set up and ready for treatment. 

Note: Saline will only flow in conjunction with ultrasound 
delivery.

-- Total Treatment Time

Number of -- 

Wounds

-- Treatment Time 

for Wound 1

PL-61044_B

       

4

Remove saline bag from outer packaging.  Remove the 

protective cover from applicator tubing spike and from 

saline bag port.  Insert spike fully into saline bag. 

9

Insert applicator tubing into the treatment wand tubing 

channel and attach the tubing clip to the treatment wand 

cable. Place treatment wand in cradle once completed. 

8

ULTRAMIST

 

SYSTEM

  

INSTRUCTIONS FOR USE

Tubing Guides

Summary of Contents for Ultramist

Page 1: ...Wand Tip APPLICATOR 18 Applicator 19 Tubing Cable Clip 20 Tubing Clamp 21 Infusion Tubing 22 Tubing Spike 23 Tubing Spike Protective Cover PLEASE READ INSTRUCTIONS BEFORE USE UltraMIST System ULTRAMIS...

Page 2: ...ed performance DESCRIPTION INDICATIONS FOR USE Description The UltraMISTTM System delivers low frequency ultrasound to the treatment site using a noncontact saline mist Indications for Use The MIST Th...

Page 3: ...the keypad to enter the size of each wound to be treated per the screen prompts 5 The total saline requirement for the combined treatment area is displayed approximately 100mL for every 25 cm2 Select...

Page 4: ...d close the pump door Leave enough tubing to allow the saline bag to hang freely with the majority of the tubing between the pump and treatment wand 13 Open the tubing clamp Saline will not flow until...

Page 5: ...or best results the treatment wand should be perpendicular to the wound and parallel to the floor during treatment Note It is important to maintain a 0 5 to 1 5 cm 0 2 0 6 inch distance from the leadi...

Page 6: ...l data is encrypted to prevent usage by unauthorized personnel WARNING DO NOT ATTACH ANY UNAPPROVED USB DRIVES CABLES OR EQUIPMENT TO THE DRIVE AS IT MAY INTERFERE WITH OPERATION OF THE DEVICE Accessi...

Page 7: ...PLICATOR Applicator has been previously used and should be replaced Replace applicator 201 REPOSITION APPLICATOR Applicator not positioned properly on treatment wand Adjust current applicator properly...

Page 8: ...software error Shut down system by turning off power and then restart If error continues call Celleration Customer Service OVERHEAT Treatment wand temperature has exceeded the allowed limit The system...

Page 9: ...r of applicator Ensure the applicator is positioned correctly with treatment wand tip located in center of applicator opening and fully seated on the applicator housing Tubing spike not fully mated wi...

Page 10: ...Celleration Customer Service PHONE 1 866 307 MIST 6478 ADDRESS 6321 Bury Drive Suite 15 Eden Prairie MN 55346 WEBSITE www Celleration com EMAIL customerservice celleration com TECHNICAL SERVICE ULTRA...

Page 11: ...re This End Up Keep Dry Recycle Authorized Representative in the European Community CE Mark Temperature Limit Consult Instructions for Use Manufacturer Date of Manufacture Lot Number Prescription Use...

Page 12: ...ation of nearby equipment It may be necessary to take mitigation measures such as re orienting or relocating the UltraMIST System or shielding the location Harmonic emissions IEC 61000 3 2 Class A Vol...

Page 13: ...stallation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful inference to radio communications Ho...

Reviews: