Caution: Federal law restricts this device to sale by or on the order 

of a licensed practitioner.

Risk of electric shock - Do not remove cover or loosen cable 

connections. Refer servicing to qualified service personnel.

Do not use if treatment wand or cables show evidence of 

deterioration or damage.  Discard disposable applicator 

components appropriately.

Risk of burns - Do not touch the metal tip of  the treatment wand 

during operation - Never depress treatment on/off button without 

applicator in place.

Always have the applicator properly seated and tubing clamp open 

when activating the treatment on/off button.

Treatment effectiveness may be diminished if any of the 

following occur:

The applicator is not fully engaged, aligned and securely 

attached to the treatment wand.

The distance between the leading edge of the applicator and 

the wound is greater than 1.5 cm (0.6 inches).

The treatment wand is not held perpendicular to the wound.

WARNINGS

PRECAUTIONS

For technical questions about the operation of the UltraMIST System contact Celleration Customer Service:
PHONE:   

1-866-307-MIST (6478)

ADDRESS:   

6321 Bury Drive, Suite 15, Eden Prairie, MN  55346

WEBSITE: www.Celleration.com
EMAIL:  

[email protected]  

TECHNICAL SERVICE

ULTRAMIST

 PRODUCT SPECIFICATIONS

SAFETY REQUIREMENTS

Meets requirements of IEC 60601-1, 3rd Edition

ETL listed as ‘Conforms to UL 60601-1’

Reference file: 3055920

ELECTRONIC COMPATIBILITY

SYSTEM CLASSIFICATION

Meets the requirements of IEC 60601-1-2 and FCC Part 18

Class II equipment

Type BF applied parts

STERILIZATION METHOD

IPX RATING

Gamma Sterilization (applicators only)

Generator: IP21

Treatment Wand: IP22

INPUT

DETACHABLE POWER CORD

CURRENT

FUSE

100-240 VAC; 50-60Hz

125V, 10A UL or equivalent listed for hospital grade.  Note: this Class II equipment 

does not rely on grounding from power cord as means of insulation.

2.0A

T2AL250V

OPERATING AND STORAGE 

CONDITIONS

Operating: Temperature: 15

º

 to 25

º C

Atmospheric Pressure: 700 to 1060 mmHg 

Applicator/System Transport and Storage: Temperature: -25

º

 to 75

º C

Relative Humidity: 30% to 75% (non-condensing)

ACOUSTIC OUTPUT POWER

Operating Frequency: 40 Khz

 

± 1 Khz

Maximum Unintentional Ultrasound Energy: < 100mW/cm

2

Tip Displacement Range: 65

µm 

±

 10µm

PL-61044_B

      

10

ULTRAMIST

 

SYSTEM

  

INSTRUCTIONS FOR USE

Summary of Contents for Ultramist

Page 1: ...Wand Tip APPLICATOR 18 Applicator 19 Tubing Cable Clip 20 Tubing Clamp 21 Infusion Tubing 22 Tubing Spike 23 Tubing Spike Protective Cover PLEASE READ INSTRUCTIONS BEFORE USE UltraMIST System ULTRAMIS...

Page 2: ...ed performance DESCRIPTION INDICATIONS FOR USE Description The UltraMISTTM System delivers low frequency ultrasound to the treatment site using a noncontact saline mist Indications for Use The MIST Th...

Page 3: ...the keypad to enter the size of each wound to be treated per the screen prompts 5 The total saline requirement for the combined treatment area is displayed approximately 100mL for every 25 cm2 Select...

Page 4: ...d close the pump door Leave enough tubing to allow the saline bag to hang freely with the majority of the tubing between the pump and treatment wand 13 Open the tubing clamp Saline will not flow until...

Page 5: ...or best results the treatment wand should be perpendicular to the wound and parallel to the floor during treatment Note It is important to maintain a 0 5 to 1 5 cm 0 2 0 6 inch distance from the leadi...

Page 6: ...l data is encrypted to prevent usage by unauthorized personnel WARNING DO NOT ATTACH ANY UNAPPROVED USB DRIVES CABLES OR EQUIPMENT TO THE DRIVE AS IT MAY INTERFERE WITH OPERATION OF THE DEVICE Accessi...

Page 7: ...PLICATOR Applicator has been previously used and should be replaced Replace applicator 201 REPOSITION APPLICATOR Applicator not positioned properly on treatment wand Adjust current applicator properly...

Page 8: ...software error Shut down system by turning off power and then restart If error continues call Celleration Customer Service OVERHEAT Treatment wand temperature has exceeded the allowed limit The system...

Page 9: ...r of applicator Ensure the applicator is positioned correctly with treatment wand tip located in center of applicator opening and fully seated on the applicator housing Tubing spike not fully mated wi...

Page 10: ...Celleration Customer Service PHONE 1 866 307 MIST 6478 ADDRESS 6321 Bury Drive Suite 15 Eden Prairie MN 55346 WEBSITE www Celleration com EMAIL customerservice celleration com TECHNICAL SERVICE ULTRA...

Page 11: ...re This End Up Keep Dry Recycle Authorized Representative in the European Community CE Mark Temperature Limit Consult Instructions for Use Manufacturer Date of Manufacture Lot Number Prescription Use...

Page 12: ...ation of nearby equipment It may be necessary to take mitigation measures such as re orienting or relocating the UltraMIST System or shielding the location Harmonic emissions IEC 61000 3 2 Class A Vol...

Page 13: ...stallation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful inference to radio communications Ho...

Reviews: