Precautions and Intended Use
3
The Air Supply MV600-2 is intended for use with the REMUV™ Air Mat and
REMUV™ LP in hospitals and long-term or extended care facilities for the later-
al transfer of patients from one bed surface to another.
INTENDED USE
The caregiver/operator is the person handling the equipment.
The patient is not the intended operator.
1. Read and understand these instruction before using this device.
2. Never leave patient unattended on an inflated device.
3. Use this product only for its intended purpose as described in this manual and the
REMUV™ or REMUV™ LP User Guide. Only use attachments and/or accessories that
are authorized by CEGA Innovations®.
4. Route the power cord to avoid hazard.
5. Avoid blocking air supply intakes.
6. When using the air supply in the MRI environment, a 25-ft. MRI hose is required
(available for purchase).
7. Avoid electric shock. Do not open the air supply. Do not attempt service or mainte-
nance while the air supply is in use.
8. NOTE: REFERENCE REMUV™ USER GUIDE FOR OPERATING INSTRUCTIONS PRIOR TO
USE.
9. Do not use the Air Supply MV600-2 in damp conditions.
10. Do not use the Air Supply MV600-2 if it is wet.
11. Wipe down the Air Supply MV600-2 with any facility approved cleaning agent only
after unplugging the product from the electrical outlet.
12. Never use the Air Supply MV600-2 if any part of the electrical cord or plug is dam-
aged. Never attempt repairs on your own. See pages 11-13 for warranty and order
information.
13. Operator must observe patient while the Air Supply MV600-2 is in use.
14. Position equipment for easy disconnection from power supply.
15. Not for use in an oxygen-enhanced environment.
16. Not for use in the presence of flammable anesthetics.
17. Avoid spillage, leakage, and ingress of liquids.
WARNING: To avoid the risk of electric shock, this equipment must only be
connected to a supply main with a protective earth ground connection.
WARNING: The Air Supply MV600-2 is not compatible with DC power
supplies.
WARNING: The Air Supply MV600-2 is intended only for use with the
REMUV™ product line.
WARNING: Risk of explosion in the presence of flammable anesthetics.
WARNING: No modification of this equipment is allowed.