Instructions 

F20B, F22B

Enter the pass code. 3510 is the factory 

default, but it is user-modifiable.
Use the 

s

 or 

t

 buttons to set the left-most 

digit to 3.

Gauge Configuration

The gauge is designed to use a 4 digit pass code to enter the 

configuration modes. This is to prevent unauthorized changing of 

settings.

With the gauge off, press and hold the 

s

 

button. Then press the Power button.

Release all buttons when the display indi-

cates 

CFG

. The gauge firmware version is 

also displayed.

hold

t

s

Zero/Clear

Memory

*

1

*

2

G

F

C

firmware version

power up sequence

_

_

_

_

CFGPC

The gauge then goes through the normal 

power up sequence.
The display prompts for entry of the configu-

ration pass code (

CFGPC

), with the first 

underscore blinking. 
Note: The gauge will automatically revert to normal operation if no 

buttons are pressed for approximately 15 seconds. To cancel and 

return to normal operation, press and release the Power button 

without entering any pass code characters.

t

s

Zero/Clear

Memory

*

*

_

_

_

3

CFGPC

- or -

_

_

5

3

CFGPC

t

s

Zero/Clear

Memory

*

_

_

_

3

CFGPC

_

I

5

3

CFGPC

0

I

5

3

CFGPC

t

s

Zero/Clear

Memory

*

0

I

5

3

CFGPC

Press and release the Power button to index 

to the next position. 
The 3 will remain, and the second position 

will be blinking.

Use the 

s

 or 

t

 buttons to select 5.

Press and release the Power  button to index 

to the next position. 3 5 will remain, and the 

third position will be blinking.
Use the 

s

 or 

t

 buttons to select 1.

Press and release the Power  button to index 

to the next position. 3 5 1 will remain, and 

the fourth position will be blinking.
Use the 

s

 or 

t

 buttons to select 0.

Press and release the Power button to pro-

ceed with configuration procedures. 
Note: If an incorrect pass code is entered, 

the gauge will return to the start of the pass 

code entry sequence.

Gauge Configuration—User or Factory

Upon successful pass code entry, the upper 

display will be blank, and the lower sec-

tion will display 

USER

. If User is not 

displayed press and release the 

t

 button 

to change the lower display to 

USER

. With 

User selected, the gauge configuration can 

be modified as described in the following 

sections. 
Press and release the Power button to con-

tinue with configuration.

F20B:

 Go to 

Engineering Unit Selection

 to 

continue user configuration.

F22B:

 Go to the 

F22B Min/Max Setup

 sec-

tion to continue user configuration.

If Factory (

FCTRY)

 is selected, the user 

configuration will be replaced by the con-

figuration as it left the factory. To select 

Factory, press and release the 

s

 button. 

The lower display will indicate 

FCTRY

.

Press and release the Power button to 

restore the factory configuration and restart the gauge.

t

s

Zero/Clear

Memory

*

USER

t

s

Zero/Clear

Memory

*

USER

t

s

Zero/Clear

Memory

*

FCTRY

*

then

- or -

Enter Configuration Pass Code

With the gauge powered up and in the 

normal operating mode, press and release 

the Memory button to sequence through the 

memory locations.

When the Memory button is pressed the 

gauge is in the peak hold mode. A new 

higher reading will replace an existing read-

ing, but a pressure reading lower then the 

one displayed will not be saved.
When desired memory location is displayed, 

take the pressure reading. The peak reading 

will be captured.
Remove the gauge from the pressure source 

and press the memory button for the next 

location.
Repeat until all readings are taken.
The readings will be saved even if the 

 

gauge is shut off.

Press and release the Power button to exit 

the memory mode and return to live pres-

sure readings.

F20B Memory

t

s

Zero/Clear

Memory

*

0

0

PSIG

Clear a Memory Location

Press and release the Memory button until 

the value to be cleared is displayed.

Press and hold the Zero/Clear button. 

Release the button when 

clr

 is displayed. 

The reading for the indicated memory loca-

tion will be cleared.
With a gauge reference models if no pres-

sure is applied, the value will return to zero.
If pressure is applied the new pressure 

reading will be stored in memory.
Absolute reference models will store the 

current atmospheric pressure reading if the 

gauge port is open to atmosphere.
Press and release the Power button to exit 

the memory mode and return to live pres-

sure readings.

0

0

 MEM I 

psig

0.0

0

0

0

2

 MEM I 

t

s

Zero/Clear

Memory

*

F22B Min/Max Memory

Press and release the Memory button to 

view the maximum stored value.
The center power button may be pressed 

at any time to return to the normal display 

mode.

The gauge may be left in the maximum dis-

play mode if desired. The maximum reading 

will be continuously displayed, stored and 

updated.

Press and release the Memory button to 

view the minimum stored value. 
For many applications it may be best to 

bring the system up to normal pressure and 

then clear the minimum value. 
The gauge may be left in the minimum dis-

play mode if desired. The minimum reading 

will be continuously displayed, stored and 

updated. 

Press and release the center power button 

to return to the normal display mode.

The Min/Max setup procedure in the 

Gauge Configuration > F22B 

Min/Max Setup

 section may be used to configure the gauge to 

capture both maximum and minimum values, the maximum value 

only, or the minimum value only. Only the configured values will 

be displayed when the memory button is pressed. The gauge also 

may be configured to erase or save the readings when the gauge 

powers down. 
The Min/Max readings are captured at the rate of 3 times per 

second. Note that if a brief pressure deviation occurs, it may not be 

captured. The readings are captured any time the gauge is on and 

not in the configuration or calibration mode.

psig

200.0

205.0

 MAX

psig

200.0

i92.0

 MIN

t

s

Zero/Clear

Memory

*

t

s

Zero/Clear

Memory

*

t

s

Zero/Clear

Memory

*

t

s

Zero/Clear

Memory

*

0

0

0

2

 MEM

0

0

 MEM I 

r

l

c

 MEM I 

t

s

Zero/Clear

Memory

*

 MIN 

 MAX 

or
or

t

s

Zero/Clear

Memory

*

Engineering Unit Selection

With the gauge in the user configuration 

mode, the upper display will be blank with 

the engineering units in the lower display.
Use the 

s

 and 

t

 buttons to navigate 

through the list of engineering units. 

Available engineering units depend on the 

sensor range.
For compound gauges the choice of CMPND 

(inHg/psig) or -/+EU (±Engineering Units) 

will apear. The gauge must be changed 

to -/+EU first before alternate engineering 

units may be selected. 
When the desired units are displayed, press 

and release the Power button to save your 

selection and move to the next parameter.

t

s

Zero/Clear

Memory

*

*

PSIG

- or -

INHG

t

s

Zero/Clear

Memory

*

Auto Shutoff Time Selection

The auto shutoff time is displayed on the 

upper display. The lower display will indicate 

AST M

 if the time displayed is in minutes 

or 

AST H

 if it is in hours. 

Use the 

s

 and 

t

 buttons to select 0 

(manual shutoff), 1, 2, 5, 10, 15, 20 or 30 

minutes, or 1, 2, 4, or 8 hours.
A setting of zero disables the auto shutoff 

timer. This requires using the Power button 

to shut the gauge off.
When the desired time is displayed, press and 

release the Power button to save your selection. 
If  the gauge was ordered with a custom 

shutoff time it will be unavailable if the time 

is changed. Reset the gauge to the original factory defaults as 

described previously to restore the custom time.

F20B:

 Go to the appropriate 

F20B Memory

 section on the next 

page to continue user configuration.

F22B:

 Go to the 

F22B Backlight Shutoff Time

 section on the next 

page to continue user configuration.

5

AST M

I

AST H

t

s

Zero/Clear

Memory

t

s

Zero/Clear

Memory

*

*

- or -

*

After the center power button is pressed 

when in user configuration mode, the dis-

play indicates 

MX/MN

.

Use the 

s

 or 

t

 buttons to select the 

desired configuration.

MX/MN

 to capture both maximum and 

minimum readings.

--/MN 

to capture minimum readings only.

MX/--

 to capture maximum readings only.

Press and release the power button to save 

the user configuration and move to the next 

setup parameter.

After the center power button is pressed 

when in user 

MX/MN 

configuration mode, 

the upper display indicates 

clr

.

Use the 

s

 button to select 

AUTO

 and the 

or 

t

 button to select 

MAN

.

When the lower display indicates 

AUTO

the maximum and/or minimum readings will 

be automatically cleared when the gauge is 

powered off. 
When the lower display indicates 

MAN

, the 

maximum and/or minimum readings will 

be retained in memory after the gauge is 

powered off. The readings can be cleared 

manually.
Press and release the power button to save 

the user configuration and move to the next 

setup parameter.

F22B Min/Max Setup

MX/--

MX/MN

--/MN

MX/--

MX/MN

MAN

r

l

c

man

- or -

*

*

r

l

c

Auto

man

t

s

Zero/Clear

Memory

t

s

Zero/Clear

Memory

*

*

- or -

t

s

Zero/Clear

Memory

*

t

s

Zero/Clear

Memory

t

s

Zero/Clear

Memory

t

s

Zero/Clear

Memory

*

cecomp.com

1220 American Way Libertyville, IL 60048 

800-942-0315

CECOMP

 

Div. of

 

A

BSOLUTE

 P

ROCESS

 I

NSTRUMENTS

Summary of Contents for F20B

Page 1: ...1 35KGCMG 01 350GCMG 1 0 1MPAG 0001 2ATMG 001 7000GCMG 1 35ATMG 01 260MMHGG 1 1BARG 001 60 psig Res 5200MMHGG 1 1000 psig Res 260TORRG 1 1KGCMG 001 60PSIG 01 5200TORRG 1 1000PSIG 1 350MBARG 1 1ATMG 001 120INHGG 1 7000MBARG 1 2040INHGG 1 350CMH2OG 1 15 psig Res 1660INH2OG 1 7000CMH2OG 1 2300FTH2O 1 3500MMH2OG 1 15PSIG 01 960ZING 1 230FTH2O 1 7000KPAG 1 12FTH2O 01 30INHG 15PSIG 01 4200GCMG 1 700KPAG...

Page 2: ... 8 2068 2041 2109 210 9 2109 210 9 48 00 6 921 83 0 155 1 155 1 6 108 0 to 5 000 psig 5 000 34 47 344 7 3447 3402 3515 351 5 3515 351 5 80 0 11 54 138 4 258 6 258 6 10 18 15 00 to 0 psi abs 15 00 abs 103 4 abs 1034 abs 1034 abs 1 034 abs 1 021 abs 1 055 abs 1055 abs 1055 abs 240 0 abs 34 61 abs 415 1 abs 775 7 abs 775 7 abs 30 54 abs 0 to 14 70 psig vac 14 70 vac 101 3 vac 1013 vac 1013 vac 1 013 ...

Page 3: ...ress and release the Memory button to view the maximum stored value The center power button may be pressed at any time to return to the normal display mode The gauge may be left in the maximum dis play mode if desired The maximum reading will be continuously displayed stored and updated Press and release the Memory button to view the minimum stored value For many applications it may be best to bri...

Page 4: ...ation Pass Code With the unit off press and hold the t button to view and or change the user calibration pass code Then press the Power but ton Release all buttons when the display indicates CAL Change Pass Code Mode Before the unit enters the view or change pass code mode the display initially indicates _ _ _ _ with the first underscore blinking and with CFGPC or CALPC on the character segments N...

Reviews: