background image

Model: THD2FE

 

NON-CONTACT FOREHEAD THERMOMETER

Forehead mode: 93.2 to 108°F/34 to 42.2°C 

Surface mode: -7.6 to +176°F/-22 to +80°C

FOR MEASUREMENT OF HUMAN BODY 

TEMPERATURE IN PEOPLE OF ALL AGES

•  Response Time: Forehead mode: 1 second;  

Surface mode: 0.6 second

•  Forehead Measurement Distance: within 1.5 in/4 cm
•  Surface mode: Distance:Spot = 1:1, emissivity: 0.95
•  Accuracy: ±0.4°F/0.2°C
•  Backlit in blue
•  Memory recalls 25 readings
•  One-button operation
•  Battery status indication
•  Temperature alert or silent
•  On/Off Power button
•  Auto-off after 1 minute
•  ABS plastic
•  Two 1.5V AAA IEC LR03 alkaline batteries, included
•  FDA, RoHS, ISO 9001, ISO 13485, ASTM E1965-98, 

EN ISO 80601-2-56

Note: 

Remove label from display before initial use.

Note:

 In the following instructions, names of the control 

buttons are shown in CAPS. Function information that 

appears on the display is shown in 

BOLD CAPS

.

USAGE

Indications for Use

The Non-Contact Forehead Thermometer is an 
infrared thermometer intended for the intermittent 
measurement of human body temperature in people 
of all ages.

Intended Operator:

At least 11 years old (5 years intensive reading 
experience), no maximum.

Ambient Temperature 

If there is any temperature difference between the 
places where the device is stored and where it will 
be used, allow the thermometer to adjust to the room 
temperature where the subject is for at least  
15 minutes before measurement.

Points of Attention 

1.  Forehead temperature is displayed in oral mode. 

This mode converts the forehead temperature to 
display its “oral-equivalent” value.

2. Before the measurement, the subject should stay 

in a stable environment for 5 minutes and avoid 
exercise or bath for 30 minutes.

3. Keep the forehead area clean and free from sweat, 

cosmetics and scars while taking the temperature.

4. The “Clinical Bias” is -2.5 to -3.1°F (-1.4 to -1.7°C).
5. The “Limits of Agreement” is 0.98.
6. The “Repeatability” is 0.36°F (0.20°C)

BEFORE WE BEGIN

This thermometer has been designed for use in 
professional healthcare facility environments and 
home healthcare environments (as defined in 3.1 of 
IEC 60601-1-11 Edition 2.0 2015-01 and 

Guidance 

for Industry and Food and Drug Administration Staff: 
Design Considerations for Devices Intended for  
Home Use

, Issued November 24, 2014). 

It is not 

meant to replace a visit to the doctor.

 Compare the 

measurement result to your regular  
body temperature. 

Consult with doctor  

if you have health concerns.

PO Box 10947  Portland, OR 97296-0947  

Tel

 800 338-5594  

Fax

 800 879-2364  

Web

 CDNkitchen.com

Reviews: