Model: THD2FE
NON-CONTACT FOREHEAD THERMOMETER
Forehead mode: 93.2 to 108°F/34 to 42.2°C
Surface mode: -7.6 to +176°F/-22 to +80°C
FOR MEASUREMENT OF HUMAN BODY
TEMPERATURE IN PEOPLE OF ALL AGES
• Response Time: Forehead mode: 1 second;
Surface mode: 0.6 second
• Forehead Measurement Distance: within 1.5 in/4 cm
• Surface mode: Distance:Spot = 1:1, emissivity: 0.95
• Accuracy: ±0.4°F/0.2°C
• Backlit in blue
• Memory recalls 25 readings
• One-button operation
• Battery status indication
• Temperature alert or silent
• On/Off Power button
• Auto-off after 1 minute
• ABS plastic
• Two 1.5V AAA IEC LR03 alkaline batteries, included
• FDA, RoHS, ISO 9001, ISO 13485, ASTM E1965-98,
EN ISO 80601-2-56
Note:
Remove label from display before initial use.
Note:
In the following instructions, names of the control
buttons are shown in CAPS. Function information that
appears on the display is shown in
BOLD CAPS
.
USAGE
Indications for Use
The Non-Contact Forehead Thermometer is an
infrared thermometer intended for the intermittent
measurement of human body temperature in people
of all ages.
Intended Operator:
At least 11 years old (5 years intensive reading
experience), no maximum.
Ambient Temperature
If there is any temperature difference between the
places where the device is stored and where it will
be used, allow the thermometer to adjust to the room
temperature where the subject is for at least
15 minutes before measurement.
Points of Attention
1. Forehead temperature is displayed in oral mode.
This mode converts the forehead temperature to
display its “oral-equivalent” value.
2. Before the measurement, the subject should stay
in a stable environment for 5 minutes and avoid
exercise or bath for 30 minutes.
3. Keep the forehead area clean and free from sweat,
cosmetics and scars while taking the temperature.
4. The “Clinical Bias” is -2.5 to -3.1°F (-1.4 to -1.7°C).
5. The “Limits of Agreement” is 0.98.
6. The “Repeatability” is 0.36°F (0.20°C)
BEFORE WE BEGIN
This thermometer has been designed for use in
professional healthcare facility environments and
home healthcare environments (as defined in 3.1 of
IEC 60601-1-11 Edition 2.0 2015-01 and
Guidance
for Industry and Food and Drug Administration Staff:
Design Considerations for Devices Intended for
Home Use
, Issued November 24, 2014).
It is not
meant to replace a visit to the doctor.
Compare the
measurement result to your regular
body temperature.
Consult with doctor
if you have health concerns.
PO Box 10947 Portland, OR 97296-0947
Tel
800 338-5594
Fax
800 879-2364
Web