TLE
F.3
OFF-SET VALUE
1) Press and hold “
TARE
”. At the same time, quickly press and release “
ON
”. Release “
TARE
”.
2) Display should read “
F.1
” (
if the display reads “
F.0
”, press “
TARE
” again, which will display “
F.1
”
).
3) To get to “
F.3
”, press “
TARE
” twice and then press “
ZERO
”
4) Display will read current value. If the unit is operating properly, this value will be between 5,000 and 15,000.
If the value is not within this range, you must adjust the off-set potentiometer on the rear PC board. If, upon
adjustment, the value remains outside the proper range, contact CCi Scale Company Inc. for instructions.
5) Return to NORMAL operation by pressing “
ON
”
F.4
AUTOMATIC SHUT OFF FUNCTION
1) Press and hold “
TARE
”. At the same time, quickly press and release “
ON
”. Release “
TARE
”.
2) Display should read “
F.1
” (
if the display reads “
F.0
”, press “
TARE
” again, which will display “
F.1
”
).
3) To get to “
F.4
”, press “
TARE
” three times and then press “
ZERO
”
4) The display will indicate either:
a) “
4 OFF
”, which indicates the unit will automatically shut itself off after four (4) minutes on non-operation.
b) “
0 OFF
”, which indicates the unit will remain in the ON mode until physically turned OFF.
5) To select the mode that is not displayed, press “
ZERO
” one time. This should change the display to the next
mode. When the display reaches your selection press “
ON
”. This returns the scale to normal operation and your
choice is implemented during future operations.
ERROR CODES
“
ERR 1
” Overloaded
The scale weighs from 0 to full capacity plus 9 divisions. The display will read “
ERR 1
” if overloaded.
“
ERR 2
” Out of zero range
Tare weight can be removed (tared off) by pressing “
ZERO
”. This range is from 0 to the full capacity plus 2
divisions. If a weight is tared and “
ZERO
” is pressed, the display will read “
ERR 2
”.
“
ERR 3
” Low battery indicator
When the display reads “
ERR 3
”, the scale must be connected to the AC adaptor and plugged into an electrical
outlet (110 V) for approximately 8 hours to fully recharge the battery.
“
ERR 4
” Dead load out of capacity
When this display appears, the dead load of platform assembly is out of tolerance (no additional mass on
scale). Call your dealer for service.
NOTE:
“
ERR 1
” and “
ERR 4
” may also indicate the
OFF-SET VALUE
is out of range. Refer to
F.3
for adjust-
ment instructions.