34
Alarms description
Alarm code
DESCRIPTION
SOLUTION
AC80 AC80-07
00
0
Microcontroller memory alarm
Contact the technician
I14
2
Amalgam level > 95%
Replace the amalgam container as
soon as possible
I15
3
Amalgam level > 100%
Replace the amalgam container
I00
32
Microcontroller memory alarm
Contact the technician
I01
33
Short-circuit due to one of the two
motors
Identify where the short-circuit is
coming from and eliminate it
I02
34
Short-circuit before the motors’ start-
ing
Contact the technician (the PC-Board
is likely to be damaged)
I03
35
Condensers have not been charged
Contact the technician (replace the
PC-Board)
I04
36
Overtemperature (temperature has
exceeded the max. allowed limit)
Ventilate the plant room
I05
37
Blower overcurrent (the blower has
exceeded the current limit)
Check the operation of the blower
(seized or excessive friction)
I07
39
Condensers overvoltage (the con-
densers have exceeded the max.
voltage limit)
Check the mains voltage (max. 260V)
S08
40
Centrifuge short-circuit
Eliminate the short-circuit
S09
41
PC-Board short-circuit on the centri-
fuge output
Replace the PC-Board
S10
42
Centrifuge instant overcurrent (the
centrifuge has exceeded the current
limit)
Eliminate siphons in the piping or
check the centrifuge operation (seized
or excessive friction)
S11
43
Centrifuge time delayed overcurrent
(the centrifuge has exceeded the
current limit)
Eliminate siphons in the piping or
check the centrifuge operation (seized
or excessive friction)
I16
48
The vacuum detecting pipe is discon-
nected
Connect the vacuum detecting pipe
to the centrifuge
Important notices
• Appliances are guaranteed for one year from the date of sale, provided that the warranty slip is
returned to the manufacturer with date of sale, retailer’s and customer’s name.
• Warranty and manufacturer’s liability cease in case appliances are treated with products which
are unsuitable or different from those recommended by the manufacturer and also in case appli-
ances are improperly used or tampered with operations of any kind carried out by people who
are not authorized by the manufacturer.
• The manufacturer, concessionaires, agents and authorized technicians are at customers’ disposal
for advice and assistance and to supply literature, spare parts and anything useful.
• The manufacturer reserves the right to modify the products for improvements, for technical, nor-
mative and functional reasons or for problems due to the availability of products or semi-fi nished
products, without prior notice.
• Our
updated
manuals are available on the web site
www.cattani.it
.
We recommend they are consulted especially for updates about
safety
.
HOME
INDEX
Summary of Contents for TURBO-SMART
Page 19: ...18 HOME INDICE ...
Page 37: ...36 HOME INDEX ...
Page 55: ...54 HOME INDEX ...
Page 73: ...72 HOME INHALTSVERZEICHNIS ...
Page 91: ...90 INDICE HOME ...
Page 99: ...HOME ...