Detergents damage aspirators, hypochlorite destroys them.

O-rings (seals) and the slides for the closure of terminals (ill. 
I) must be lubricated with Lubri-Jet every 15 working days. 
It is important to replace all fl exible tubes once a year (ill. 
L): in particular, the external hose and terminals (ill. L) must 
be disinfected for each patient and replaced every three 
months for hygienic and operational reasons.

• After every surgical operation or very long treatment: rin-
se the appliance by aspirating water, preferably warm and 
Puli-Jet plus new. Replace and disinfect external tubes and 
terminals.

• After every working day: clean the fi lters; clean the equi-
pment with warm water and Puli-Jet plus new. Insert the 
disinfectant antifoam tablets in the fi lter.

• Every 15 days: clean the canister, the draining valve and 
probes; lubricate O-rings and slides closing terminals with 
Lubri-Jet.

• Periodical checks: the service engineer shall check up all 
relays, functioning of probes and draining pump, siphons 
and waste outlets, all internal pipes, and plastic and rubber 
compo-nents subject to ageing; hoses and seals should be 
replaced once a year. The appliance must be tested periodi-
cally from the current safety regulations point of view. Short-
circuit probes: make sure that suction stops and that the 
drainage pump starts operating. 

• The manufacturer is at your disposal to supply spare parts, 
literature, instructions and anything that might be useful. 
Our concessionaires, agents, authorised retailers and servi-
ce engineers are provided with split-up drawings, electrical 
diagrams, instructions and updating about maintenance and 
service.

• The appliance is guaranteed for a year from the date of 
sale provided that the guarantee card addressed to the ma-
nufacturer is returned with date of sale, retailer’s stamp and 
customer’s name.

Summary of main cleaning and routine 

maintenance operations to be entrusted to 

purposely trained surgery staff

Extraordinary maintenance operations to 

be entrusted to trained service engineers, 

provided with original spare parts

14

• 

Summary of main cleaning and routine maintenance  

 

 

operations to be entrusted to purposely trained  

 

 

 surgery 

staff

• 

Extraordinary maintenance operations to be entrusted  

 

 

to trained service engineers, provided with original spare parts

Ill. H

Ill. I

Ill. L

Summary of Contents for Mono-Jet a

Page 1: ...MANUALE ISTRUZIONI OPERATOR S HANDBOOK MANUEL D UTILISATION GEBRAUCHSANWEISUNG MANUAL DE ISTRUCCIONES MONO JET...

Page 2: ...nce and cleaning 13 Summary of main cleaning and routine maintenance operations to be entrusted to pur posely trained surgery staff 14 Extraordinary maintenance operations to be entrusted to trained s...

Page 3: ...02 Off IEC 417 5008 On IEC 417 5007 This appliance cannot work in the presence of an anaesthetic mixture inflammable with air of oxygen or of nitrogen protoxide Thermal motor protector Power output 0...

Page 4: ...e box General features Installation The best position for the Mono Jet is near the dental unit precisely on the left of the chair close to the patient s feet ill A The 3b tube holder must be connected...

Page 5: ...logical filter is available to filter the exhausted air ill C pag 4 Liquids heavier than air are collected on the bottom of the canister Draining is automatic the valve 6 ill D placed on the bottom of...

Page 6: ...n for some seconds with open terminals aspirating air only so that hoses and manifold dry completely inside then cut off the power and pull out the filter To inspect the filter 9 lift up the pipe hold...

Page 7: ...r compo nents subject to ageing hoses and seals should be replaced once a year The appliance must be tested periodi cally from the current safety regulations point of view Short circuit probes make su...

Page 8: ...appliances and or the plants have been tampered with operations of any kind done by persons unfit and so unauthorised by the manufacturer For any application that is not included and described in this...

Page 9: ...NTERR PORTELLO DOOR SWITCH S 1 INTERR GENERALE MASTER SWITCH TR 4VA T 1 BT 150 P 1 WL 02 C 2 47 mF C 1 0 1mF R 2 1K R 1 270 RS RELE SONDE PROBES RELAY RP RELE POMPA PUMP RELAY RM RELE MOTORE 14 06 96...

Page 10: ...7 Gruppo aspirante Uni Jet 75 10 Manicotto portatubo secreti sul vaso portasecreti 11 Vaso portasecreti Maxi Canister 12 Pompa di drenaggio LEGEND OF COMPONENTS 1 On Off switch 2 Exhausted air silence...

Page 11: ...58 FR 222 394 395 PENDING PATENT CATTANI IT M098A000019 IT M098A000119 EP 99830010 7 EP 99830011 5 EP 99830250 9 EP 00830491 7 IT M099A000165 US 09 624 182 6 A Via Natta 43122 Parma Italy 10 21 48 EST...

Reviews: