– 55 –
IT-9000
[3] Current - Printer
[Preparation]
(1) Open the battery pack cover and remove the battery pack.
(2) Connect equipment as shown below.
(3) Set the voltage control of the DC power supply at a minimum position, and then turn ON its
power.
Diode
(IF=3.0 A Max.)
DC power supply unit
DC voltmeter
Turn the right lock switch for the battery pack cover
to the “LOCK” position.
DC ammeter
No.
Operation
Display
Test Details
1
Adjust the DC power supply
output so that the DC
voltmeter reads 7.4 V, and
then turn ON IT-9000.
Select “[3].Current - Printer”
from MENU-8 screen.
Automatically judges if the
current value is within 2300 to
2900 mA.
Then, manual judgment screen
appears.
If OK.
Press the “1” key to return to
the menu screen.
If out of range (Failure):
Press the “0” key to return to
the menu screen.