Figure 22 Calibrate Screen
I. ADJUSTMENT OF LOW SET POINT
(Fig. 23)
Use UP/DOWN keys to alter the low set point in l/min. Press ENTER to continue. The pump
will start.
Figure 23 Low Set Screen
J. CALIBRATION OF LOW SET POINT
(Fig. 24)
Check the external calibrator reading equals the low set point value. Use UP/DOWN keys to
adjust the flow rate until they are equal. Press ENTER to save the new setting and continue.
Figure 24 (Low Set) Adjust Screen
K. ADJUSTMENT OF HIGH SET POINT
(Fig. 25)
Use UP/DOWN keys to alter the high set point in l/min. Press ENTER to continue.
Figure 25 High Set Screen
L. CALIBRATION OF HIGH SET POINT
(Fig. 26)
Check the external calibrator reading equals the high set point value. Use UP/DOWN keys
to adjust the flow rate until they are equal. Press ENTER to save the new setting and
continue.
The pump will stop and the unit will default to the
MAIN MENU
.
Figure 26 (High Set) Adjust Screen