45 |
P a g e
UNIT SPECIFICATIONS
Recovery Time:
From a 30-second door opening.
Model
Recovery to 80°C
Recovery to 150°
Recovery to 306°C
TFO-1
1.5 Minutes
4.5 Minutes
14.0 Minutes
TFO-3
1.5 Minutes
2.0 Minutes
18.0 Minutes
TFO-5
2.0 Minutes
4.0 Minutes
45.0 Minutes
Recovery Time:
From a 60-second door opening.
Model
Recovery to 80°C
Recovery to 150°
Recovery to 306°C
TFO-1
4.0 Minutes
6.0 Minutes
20.0 Minutes
TFO-3
2.0 Minutes
4.5 Minutes
36.0 Minutes
TFO-5
3.5 Minutes
6.0 Minutes
57.0 Minutes
A
IR
F
LOW
P
ERFORMANCE
Ventilation Rates
Model
Cubic Feet per Minute @80°C
Liters per Minute @80°C
TFO-1
5.6
158.6
TFO-3
11.3
320.0
TFO-5
15.2
430.4
Air Changes per Hour
Model
@80°C
TFO-1
330
TFO-3
225
TFO-5
180
Air Velocity Across Shelf Space
Model
Linear Feet per Minute
Meters per Minute
TFO-1
125
38.1
TFO-3
140
42.7
TFO-5
145
44.2