OPERATOR MAINTENANCE
Warning
: Disconnect the unit from its power supply prior to performing maintenance or services.
Avertissement
: Avant d'effectuer toute maintenance ou entretien de cet appareil, débrancher le cordon
secteur de la source d'alimentation.
C
LEANING
If a hazardous material or substance has spilled in the unit, immediately initiate your site’s Hazardous
Material Spill Containment protocol. Contact your local Site Safety Officer and follow instructions per
the site policy and procedures.
•
The unit chamber should be cleaned prior to first use.
•
Periodic cleaning is required.
•
Do not use spray on cleaners or disinfectants. These can leak through openings and
coat electrical components.
•
Consult with the manufacturer or their agent if you have any doubts about the
compatibility of decontamination or cleaning agents with the parts of the equipment or
with the material contained in it.
•
Do not use cleaners or disinfectants that contain solvents capable of harming paint
coatings or stainless steel surfaces.
Do not use chlorine-based bleaches or abrasives;
these will damage the chamber liner.
Warning
: Exercise caution if cleaning the unit with alcohol or flammable cleaners. Always allow the
unit to cool down to room temperature prior to cleaning and make sure all cleaning agents have
evaporated or otherwise been completely removed prior to putting the unit back into service.
Avertissement:
Soyez prudent lorsque vous nettoyez l'appareil avec de l'alcool ou des produits de
nettoyage inflammables. Laissez toujours refroidir l'appareil à la température ambiante avant le
nettoyage et assurez-vous que tous les produits de nettoyage se sont évaporés ou ont été
complètement enlevés avant de remettre l'appareil en service.
Cleaning
1.
Remove all removable interior components such as shelving and accessories.
2.
Clean the unit with a mild soap and water solution, including all corners.
Do not
use an
abrasive cleaner that will damage metal surfaces.
Do not use deionized water to rinse or
clean with.
3.
Rinse with distilled water and wipe dry with a soft cloth.
4.
Take special care when cleaning around the temperature sensor probes in the chamber to
prevent damage. Do not clean the probes.