MAXTRAP 50 & 105 Supplement Instruction Manual Rev 11.29.18

www.cascadesciences.com | [email protected] | 503-847-9047

Operation Overview

Start Up

Remove the COLDWELL Stainless jar.  Make sure your have 800 ml of a heat transfer fluid such as 

ethanol, glycol or a cryocool inside the reservoir.  

Do not pour warm fluid into unit at any time as it may cause a refrigerant over pressure situation, 

rupture potential and will void warranty or repair.

Power the MAXTRAP “ON.” The power switch is located on the right side of the Labconco unit.  

Feel the right side to make sure air is drawing in thru the side vent.  If connected to a vacuum oven, it 

does not have to be under vacuum to operate. 

Allow at least 30 minutes for the MAXTRAP fluid to reach operating temperature. 

From The Vacuum Oven:

• 

Vacuum valve is closed.

• 

Vent Valve is closed.

• 

Product is loaded inside oven.

• 

Energize the Vacuum Pump

• 

Open vacuum valve

• 

Vacuum valve closed at oven. 

• 

Vacuum pump off. 

• 

Allow unit to naturally thaw to ambient prior to opening/collection of volatiles. 

• 

To remove lid on COLDWELL, loosen/remove clamp and pull lid upwards being careful not 

to bump the baffling attached to lid unit or snag on viton gasket. 

• 

Oil rotary vane pump precaution. If you are planning to collect, reclaim, re-use the 

condensates in the trap strongly consider a FORELINE TRAP between an oil rotary vane pump 

and the MAXTRAP. Contact Cascade Sciences to purchase a trap.

Shut Down

As vapors offgas inside the oven they will pass thru the trap and collect inside the COLDWELL.

Summary of Contents for MAXTRAP 105

Page 1: ...s MAXTRAP 50 105 Supplement Instruction Manual Rev 11 29 18 www cascadesciences com info cascadesciences com 503 847 9047 Supplemental Installation Operation Manual MAXTRAP 50 105 Refrigerated Cold Tr...

Page 2: ...protection impairment by disabling or interfering with unit safety features This can result in damage to the unit and injury death or negative effects Follow all national laws regulations and local o...

Page 3: ...llowing information ready when calling or emailing Technical Support The model number and the serial number 503 847 9047 info cascadesciences com Cascade Sciences 6725 NE Evergreen Pkwy 106 Hillsboro...

Page 4: ...e period otherwise Cascade Sciences will not be held responsible Take photos of any shipping damage Refuse delivery or make note on the bill of lading 1 If anything is missing please call us at 503 84...

Page 5: ...separately from the MAXTRAP It is strongly recommend that the vacuum pumps GAS BALLAST be in the open position to best tolerate ethanol and other gas loads In some cases the vacuum pump can be powered...

Page 6: ...2 on the right side of the unit for venti lation and air to draw thru the slotted right side of the Thermo unit Electrical 120V The Labconco refrigerated base unit requires a standard 110 120 volt NEM...

Page 7: ...OVEN Make sure your have 800 ml of a heat transfer fluid such as ethanol glycol or a cryocool inside the thermo reservoir This fluid is sold separately 1 Remove the stainless COLDWELL Pour the 800ml o...

Page 8: ...l as important notes in the user manual Symbol Definition Consult the user manual Consulter le manuel d utilisation AC Power Rep re le courant alternatif AC Power Rep re le courant alternatif Begin he...

Page 9: ...not have to be under vacuum to operate Allow at least 30 minutes for the MAXTRAP fluid to reach operating temperature From The Vacuum Oven Vacuum valve is closed Vent Valve is closed Product is loade...

Page 10: ...ing Precautions 1 Do not move during operation 2 SAFETY TIP At very cold temperatures solvents and even oxygen can re condense back into liquid This is potentially explosive Especially if the cold tra...

Page 11: ...s that contain solvents capable of harming paint coatings or stainless steel surfaces Do not use chlorine based bleaches or abrasives these will damage the COLDWELL liner Vacuum Pump Maintenance Vacuu...

Reviews: