8
Service Manual –
PB Series
REV NO
:
2
4.1.2. SPAN Calibration Setting (C-3)
(1)
Pressing and holding
“
Calibration Switch
”
press [POWER] key.
After
“
CAL
”
message blinks three times and shows the version of scale, it displays
“
CAL 1
”
message.
(2)
Press [ZERO] to display
“
CAL-3
”
.
(3)
Press [TARE] key and then it displays
“
zero
”
message.
(4)
Press [TARE] key and then it displays
“
midup
”
message
(5)
Load middle weight (ex:1/3 full capacity) on the platform
(6)
Press [TARE] key and then it displays
“
span
”
message
(7)
Load full weight on the platform
(8)
Press [TARE] key and then it displays
“
middn
”
message
(9)
Load middle weight (ex:1/3 full capacity) on the platform
(10)
Press [TARE] key and then it display
“
CAL 3
”
message
4.1.3. Gravity Constant Value Setting (C-9)
Current gravitational Acceleration value is set to 9.7994 m/s
2
.
(11)
Pressing and holding “Calibration Switch” press [POWER] key.
After “CAL” message blinks three times and shows the version of scale, it displays “CAL-1”
message.
(12)
Press [ZERO] to display “C-9”.
(13)
Press [TARE] key, and then “ G-1“ message and “9.7994” will be shown. The first
digit,”9” will blink.
(14)
Input a gravitational acceleration value by using [ZERO] key.
(15)
Press [TARE] key, and then “G-2“ message blinks.“9.7994” will be shown. The first
digit,”9” will blink.
(16)
Input a gravitational acceleration value by using [ZERO] key.
(17)
Press [TARE] key to save the gravitational acceleration value, and “C-9 ” message will be
shown.
4.1.4. Calibration factor Setting (C-10)
(1)
Pressing and holding “Calibration Switch” press [POWER] key.
(2)
After “CAL” message blinks three times and shows the version of scale, it displays “CAL-1”
message.
(3)
Press [ZERO] to display “C-10”.
(4)
Press [TARE] key, and then “UNIT “ message and “0” will be shown. The first digit,”0”
will blink. It means calibration unit is “kg” (0 : kg, 1 : lb)
(5)
Input a calibration unit by using [ZERO] key.
(6)
Press [TARE] key, and then “CAPA“ message blinks.“0015” will be shown. The first digit,”0”
will blink. It means a full-capability is “15 (calibration unit, kg or lb)”
(7)
Input a capability by using [ZERO] key.
Summary of Contents for PB Series
Page 16: ...16 Service Manual PB Series REV NO 2 8 1 2 Scale Ass y ...
Page 17: ...17 Service Manual PB Series REV NO 2 8 2 Loadcell Ass y ...
Page 18: ...18 Service Manual PB Series REV NO 2 8 3 Main PCB Ass y Top 8 4 Main PCB Ass y Bottom ...
Page 19: ...19 Service Manual PB Series REV NO 2 8 4 1 Main PCB Part List ...