18
/
44
Zero /Tare key Operation mode selection
F07
●
0
Activate when “Steady” condition, only
1
Always activated
Zero key Operation Range selection
F08
0
Activated within 2% of Max. Capacity
1
Activated within 5% of Max. Capacity
●
2
Activated within 10% of Max. Capacity
3
Activated within 20% of Max. Capacity
4
Activated within 100% of Max. Capacity
Tare key Operation Range selection
F09
0
Activated within 10% of Max. Capacity
1
Activated within 20% of Max. Capacity
●
2
Activated within 50% of Max. Capacity
3
Activated within 100% of Max. Capacity
“Hold” Mode selection
F10
●
0
Peak Hold : Measure Max. weight value and hold on display.
1
Sample Hold : Hold current weight until “Hold Reset”.
2
Average Hold : Make average during 8sec, and hold display
External Input Selection
F11
Set Value
Input 1
Input 2
Input 3
Input 4
0
Zero
Tare/Reset
Hold/Reset
●
1
Run/Stop
Tare/Reset
Zero
2
Run
Stop
Sub-Total
Grand-Total
3
Zero
Tare
Tare Reset
4
Run
Stop
Tare
Tare Reset
5
Run
Stop
Zero
Code No. setting
F12
●
0
Fixed Code No.
1
Increase Code No., whenever finish one weighing process
2
Decrease Code No., whenever finish one weighing process