29
16. CHECK MESSAGE
(1) Operating Mode
CH 01
Reason
▣
Internal RAM is erased.
Solution
☞
Confirm the setting value in SET Menu.
CH 02
Reason
▣
Load cell connection failure or error in A/D conversion part.
☞
Solution
Check the Load Cell connector to see if the polarity of signal is reversed.
CH 03
Reason
▣
stable to initialize the scale.
Solution
☞
Place the scale on a stable surface.
CH 04
Reason
▣
Zero range deviate from the set range (±10% of Max.)
Solution
☞
Confirm that there is nothing on the weighting platform.
Over
Reason
▣
The display weight is larger than the Maximum Capacity you've set.
Solution
☞
Don't load the article whose weight is larger than the Max. Capacity on the
platform scale. This may damage Load Cell.