21
CAL 5
(Direct Calibration)
CAL 5-1
Function: Direct input about the zero value of loadcell
Range of set value: 1 ~ 99,999
KEY
DISPLAY
DESCRIPTION
ZE-CAL
Direct zero input mode
0.0000
Loadcell zero = 0.0000 mV/V
▲
Key : Value change
◀
Key : Digit change
ZERO Key : Cancel
SET Key : Save and next
menu
0.1000
Loadcell zero = 0.1000 mV/V
CAL 5-2
Function: Direct input about the span value of loadcell
Range of set value: 1 ~ 99,999
KEY
DISPLAY
DESCRIPTION
SP-CAL
Direct span input mode
1.2000
Loadcell span = 1.2000 mV/V
▲
Key : Value change
◀
Key : Digit change
ZERO Key : Cancel
SET Key : Save and next
menu
2.0000
Loadcell span = 2.0000 mV/V
Summary of Contents for CI-150A
Page 1: ...2 ...
Page 3: ...4 ...
Page 4: ...5 ...
Page 8: ...9 MEASURE OF APPEARANCE ...
Page 66: ...67 MEMO ...
Page 67: ...68 MEMO ...
Page 68: ...69 MEMO ...
Page 69: ...70 ...