MODEL 7545-2 

INST-PM-7545-2-UVA 

Summary of Contents for 7545-2

Page 1: ...MODEL 7545 2 INST PM 7545 2 UVA ...

Page 2: ...Revision Log Revisions Description Letter Date A 04 16 2018 Initial release ...

Page 3: ...ntenance Kit for Model 7545 2 UVA are shown below Note Photos may have a different appearance depending on chamber options 1 Turn off chamber and unplug power cord INSTRUCTIONS FOR REPLACING AIR FILTER ANTIMICROBIAL COPPER ROD AND UVA LAMPS IN MODEL 7545 2 REPLACING AIR FILTER Disconnect electrical power before opening lower grill assembly Lamps Velcro Discs Air Filter Antimicrobial Copper Rod ...

Page 4: ...pulling away from chamber to release 2 bottom latches as shown below 4 Unfasten 4 Velcro connections on air filter and discard filter as shown below Note Remove and discard any Velcro discs with loops that may still be attached to hook discs located on condenser and replace hook discs as needed Lower Grill Assembly Screws Lower Grill Assembly Latches Air Filter Velcro Connection Locations ...

Page 5: ...enter new air filter with condenser edges as shown below and press into place with cardboard side towards condenser Note Once adhesive has touched air filter it cannot be adjusted 7 Reinstall lower grill assembly by engaging 2 bottom latches 8 Secure lower grill assembly with retained 2 screws using Phillips screwdriver bit Velcro Disc Locations Air Filter ...

Page 6: ...have a different appearance depending on chamber options 1 Slide top light bank shelf out of chamber as shown below 3 Remove light bank shelf and place lamp side up on flat surface as shown below REPLACING LAMPS Top Light Bank Shelf Top Light Bank Shelf ...

Page 7: ...ng out of lamp sockets as shown below Repeat for each lamp to be replaced 5 Insert new lamp ends into lamp sockets and rotate turn Repeat for each lamp to be replaced 6 Slide light bank shelf into chamber and press firmly to seat rear connector as shown below Lamp Socket Light Bank Shelf Press Here ...

Page 8: ...g on chamber options 1 Remove bottom shelf as shown below 2 Remove 6 screws securing access panel using Phillips screwdriver bit as shown below Note Retain Screws 3 Remove access panel as shown below REPLACING ANTIMICROBIAL COPPER ROD Bottom Shelf Access Panel Screw Locations Access Panel ...

Page 9: ...nto clip Note Clip must be oriented as shown below after installing rod 7 Reinstall humidity diffuser with antimicrobial copper rod towards front of chamber 8 Reinstall access panel 9 Secure access panel with retained 6 screws using Phillips screwdriver bit 10 Reinstall bottom shelf 11 Continue to instructions for replacing additional accessory preventative maintenance items if applicable or plug ...

Page 10: ...ressing Settings button on right side of Home screen as shown below 13 Press Maint button as shown below 14 Press Reset buttons for Replace air filter and Replace UVA lamps as shown below 15 Press Home button as shown below Settings Button Reset Buttons Home Button Maint Button ...