Installing Drain Water Pump (PUMP301)
In applications where a floor drain is not available and a CARON water recycling system
is not being used, a drain pump can be purchased to pump any excess condensate from
the chamber to a local sink or drain. The pump is located near the middle of the back of
the chamber. Connect the supplied tubing from the pump to the sink / drain. The
tubing may be run vertically into a ceiling but should not exceed 15 feet height. The
pump is equipped with a small reservoir on the bottom of the pump with an internal level
switch that will automatically turn the pump
ON
when it is full to drain the water out of the
reservoir and into a floor or sink drain.
Pump Outlet
to Sink or Floor
Drain
Pump Inlet
from Chamber
Drain
Reservoir with
Internal Level
Switch