Safety and Regulatory Information
AA5348 | 2012-12-20
1–7
Note
U
T
is the mains (ac) voltage prior to application of the test level.
Electrical fast transient/
burst
IEC 61000–4–4
± 2 kV for
power supply
lines
± 1 kV for
input/output
lines
± 2 kV for
power supply
lines
± 1 kV for input/
output lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000–4–5
± 1 kV line to
line
± 2 kV line to
earth
± 1 kV line to
line
± 2 kV line to
earth
Voltage dips, short
interruptions, and volt-
age variations on
power supply lines
IEC 61000–4–11
< 5 % U
T
(> 95 % dip
in U
T
) for 0.5
cycle
40 % U
T
(60 %
dip in U
T
) for 5
cycles
70 % U
T
(30 %
dip in U
T
) for
25 cycles
< 5 % U
T
(> 95 % dip in
U
T
) for 5 sec
< 5 % U
T
(> 95 % dip
in U
T
) for 0.5
cycle
40 % U
T
(60 %
dip in U
T
) for 5
cycles
70 % U
T
(30 %
dip in U
T
) for 25
cycles
< 5 % U
T
(> 95 % dip in
U
T
) for 5 sec
Power frequency
(50/60 Hz) magnetic
field
IEC 61000–4–8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commer-
cial or hospital environment.
Electromagnetic Immunity for Equipment and Detectors Fully Compliant with IEC 60601–1–2:2007