Safety and Regulatory Information
1–6
AA5348 | 2012-12-20
Recommended EMC Environment
Guidance and Manufacturer’s Declaration—Electromagnetic Emissions
The detector is intended for use in the electromagnetic environment specified below. The customer or
the user of the detector should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment—Guidance
RF Emissions
CISPR 11
Group 1
The detector uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class A
The detector is suitable for use in all establishments other
than domestic and those directly connected to the public
low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonics Emissions
IEC 61000–3–2
Class A
Voltage
Fluctuations/
Flicker Emissions
IEC 61000–3–3
Complies
Electromagnetic Immunity for Equipment and Detectors Fully Compliant with IEC 60601–1–2:2007
The detector is intended for use in the electromagnetic environment specified below. The customer or
the user of the detector should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment—Guidance
Electrostatic Discharge
(ESD)
IEC 61000–4–2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30 %.