28 Chapter 5 Acquiring Images using the Cephalometric Modality
Acquiring a Frontal AP and PA Image for the Adult and Pediatric
Patient
Before acquiring an image, check that you have:
•
Selected the patient record.
•
Accessed the CS Imaging Software.
•
Accessed the
Cephalometric Acquisition
Interface.
Preparing the Unit and Setting the Acquisition Parameters
To set the acquisition parameters, follow these steps:
1.
On the unit, position the head clamps manually for the frontal exam.
The frontal AP exam option becomes active
on
the
Cephalometric Acquisition Interface.
2.
Select the PA exam option if you require it.
3.
S
elect the appropriate FoV option.
Important: You must position the head clamps manually because they
cannot be positioned automatically from the Program pane exam
options. When you position the head clamps manually, the relevant
exam option icon becomes active.