Carestream CS 8100 User Manual Download Page 10

6

Chapter 2  Cephalometric Modality Overview

Figure 4

CS 8100SC 3D Functional Components

1

Unit rotative arm

11

Unit column 

2

Unit head

12

Frankfort positioning indicator

3

Emergency stop knob

13

Secondary collimator

4

ON/OFF button

14

Nasion support

5

Cephalostat arm

15

Head clamps and ear cones

6

Cephalostat head

16

Cephalometric digital sensor

7

X-ray source assembly

17

X-ray remote control

8

Head and chin rest

18

Carpus panel (optional)

9

Chin rest and bite block

19

Computer hosting the imaging and the 
acquisition software

10

Panoramic/3D digital sensor

17

19

4

12

5

13

3

14

15

16

7

8

9

10

1

11

6

2

18

Summary of Contents for CS 8100

Page 1: ...User Guide CS 8100 CS 8100 3D Family CS 8100 CS 8100 Access CS 8100SC CS 8100SC Access CS 8100SC 3D CS 8100SC 3D Access...

Page 2: ...focused teeth acquisition The system has the capability to expand the field of view to the 3D full upper and lower jaw acquisition with license upgrade CS 8100SC 3D panoramic modality dental volumetri...

Page 3: ...tching on the Unit 19 Increasing the operating life of the X ray tube 19 Firewall or Network Settings 20 Accessing the Cephalometric Acquisition Interface 21 Chapter 5 Acquiring Images using the Cepha...

Page 4: ...Carpus Image for the Adult and Pediatric Patient available as an option 43 Preparing the Unit and Setting the Acquisition Parameters 43 Preparing and Positioning the Adult and Pediatric Patient 44 Lau...

Page 5: ...ize information or indicate potential risk to personnel or equipment WARNING Warns you to avoid injury to yourself or others by following the safety instructions precisely CAUTION Alerts you to a cond...

Page 6: ...2 Chapter 1 Conventions in This Guide...

Page 7: ...per and lower jaw acquisition CS 8100 3D Access panoramic modality and dental volumetric reconstruction modality limited to 3D focused teeth acquisition The system has the capability to expand the fie...

Page 8: ...nd down movement of the unit mobile components Rotation and translation movement of the rotative arm and the cephalostat Figure 1 CS 8100SC Unit Mobile Components Figure 2 CS 8100SC 3D Unit Mobile Com...

Page 9: ...it head 12 Frankfort positioning indicator 3 Emergency stop knob 13 Secondary collimator 4 ON OFF button 14 Nasion support 5 Cephalostat arm 15 Head clamps and ear cones 6 Cephalostat head 16 Cephalom...

Page 10: ...collimator 4 ON OFF button 14 Nasion support 5 Cephalostat arm 15 Head clamps and ear cones 6 Cephalostat head 16 Cephalometric digital sensor 7 X ray source assembly 17 X ray remote control 8 Head a...

Page 11: ...Ed02 7 Digital Sensor Locations The following figures illustrates the locations of the digital sensors and cepahometric sensors of the units Figure 5 CS 8100SC digital sensor and cephalometric sensor...

Page 12: ...al components of the CS 8100 Family head and chin rest Figure 7 CS 8100 Family Head and Chin Rest Functional Components 1 Positioning Panel 5 Bite block support 2 Frontal head rest adjustment knob 6 C...

Page 13: ...Head and Chin Rest Figure 8 illustrates the functional components of the CS 8100 3D Family head and chin rest Figure 8 CS 8100 3D Family Head and Chin Rest Functional Components 1 Positioning Panel 4...

Page 14: ...s the rotative arm in a parallel position to the unit head leaving more free space around the unit To do so press and hold both buttons until the Ready Indicator LED switches off When you release both...

Page 15: ...etric Modality Positioning Accessories The following accessories are used when positioning a patient using the Cephalometric Modality Table 1 Cephalometric Modality Positioning Accessories Accessory D...

Page 16: ...utside the X ray room You must press and hold the exposure button until the end of acquisition Premature release of the exposure button interrupts the acquisition Figure 11 X ray Remote Control 1 Expo...

Page 17: ...igned and developed specifically to support image diagnosis Acquisition Interface The Acquisition interface is a user friendly interface specifically designed and developed for the CS 8100 and CS 8100...

Page 18: ...fore the next acquisition 6 X ray dose Displays the dose that will be emitted before each acquisition 7 Reset button Resets the unit to the patient entry position to enable you to position a patient i...

Page 19: ...al exam options Click for a lateral exam Click for an oblique exam Click for a frontal AP Anterior Posterior exam Click for a submento vertex exam Click for a frontal PA Posterior Anterior exam Click...

Page 20: ...eters must be based on the age and morphology of the patient Figure 3 Cephalometric Acquisition Interface Patient Pane Note A selected patient mode sets the kV and mA Patient type parameters Pediatric...

Page 21: ...adiological image acquisition If the default parameter setting is not adapted to your patient type you can manually adapt the parameter setting to the patient type and save this setting as the default...

Page 22: ...18 Chapter 3 Imaging Software Overview...

Page 23: ...e will close 3 You can now proceed to start the Acquisition interface from the imaging software Increasing the operating life of the X ray tube To increase the operating life of the X ray tube follow...

Page 24: ...s follow these steps 1 On your desktop double click to open the CS Imaging Software The Windows Security Alert dialog box displays Select Private networks Public networks and click Allow access 2 Clic...

Page 25: ...tion Interface follow these steps 1 Select and double click the patient card in the patient list The Imaging window is displayed 2 In the Imaging window click to access the Cephalometric Acquisition I...

Page 26: ...22 Chapter 4 Getting Started...

Page 27: ...sition Interface Preparing the Unit and Setting the Acquisition Parameters To set the acquisition parameters carry out the following steps 1 On the unit position the head clamps manually for the later...

Page 28: ...your patient type click the Parameter button Select the appropriate parameters Click and select Memorize Anatomy setting if you want to save the new parameter settings as future default settings Impo...

Page 29: ...nt to wear a lead apron Ensure that the apron lays flat across the patient s shoulders 3 Pull out at the top part of a head clamp bar to extend both bars 4 Ask the patient to do the following Move for...

Page 30: ...clamp bar to retract both head clamp bars and to fit the ear cones into the ear auditory canals 7 Lower the nasion support to a vertical position 8 Use the Frankfort positioning indicator as a visual...

Page 31: ...ed to the CS Imaging Software 3 Check the image quality if satisfactory remove the ear cones and the nasion support and release the patient X ray Dose Emission Information Compliance with EURATOM 97 4...

Page 32: ...sition Parameters To set the acquisition parameters follow these steps 1 On the unit position the head clamps manually for the frontal exam The frontal AP exam option becomes active on the Cephalometr...

Page 33: ...ot adapted to your patient type click the Parameter button Select the appropriate parameters Click and select Memorize Anatomy setting if you want to save the new parameter settings as future default...

Page 34: ...s the patient s shoulders 3 Pull out at the top part of a head clamp bar to extend both bars 4 Ask the patient to do the following Move forward Stand up straight in the following positions For a front...

Page 35: ...S 8100 3D Family_SM987_Ed02 31 6 Gently push in at the top part of a head clamp bar to retract both head clamp bars and to fit the ear cones into the ear auditory canals Figure 5 1 Frontal AP Figure 5...

Page 36: ...steps 1 Leave the X ray room and close the door You must keep visual contact with the patient during acquisition Figure 6 Frontal AP Figure 7 Frontal PA Important If you have a problem that requires y...

Page 37: ...d to the CS Imaging Software 3 Check the image quality if satisfactory remove the ear cones and the nasion support and release the patient X ray Dose Emission Information Compliance with EURATOM 97 43...

Page 38: ...sition parameters follow these steps 1 On the unit position the head clamps manually for an oblique exam The oblique exam option becomes active 2 On the Cephalometric Acquisition Interface select the...

Page 39: ...1 Ask the patient to remove all metal objects 2 Ask the patient to wear a lead apron Ensure that the apron lays flat across the patient s shoulders 3 Pull out at the top part of a head clamp bar to e...

Page 40: ...bars and to fit the ear cones into the ear auditory canals 7 Lower the nasion support to a vertical position Launching the X ray To launch the X ray follow these steps 1 Leave the X ray room and close...

Page 41: ...d to the CS Imaging Software 3 Check the image quality if satisfactory remove the ear cones and the nasion support and release the patient X ray Dose Emission Information Compliance with EURATOM 97 43...

Page 42: ...t the acquisition parameters follow these steps 1 On the unit position the head clamps manually for a frontal AP exam The frontal AP exam option becomes active 2 On the Cephalometric Acquisition Inter...

Page 43: ...patient follow these steps 1 Ask the patient to remove all metal objects 2 Ask the patient to wear a lead apron Ensure that the apron lays flat across the patient s shoulders 3 Pull out at the top par...

Page 44: ...etric Modality 5 On the control panel press and hold to level the ear cones to the patient s auditory canals 6 Gently push in at the top part of a head clamp bar to retract both head clamp bars and to...

Page 45: ...To launch the X ray follow these steps 1 Leave the X ray room and close the door You must keep visual contact with the patient during acquisition Important If you have a problem that requires you to s...

Page 46: ...maging Software 3 Check the image quality if satisfactory remove the ear cones and the nasion support and release the patient X ray Dose Emission Information Compliance with EURATOM 97 43 Directive Yo...

Page 47: ...ion the head clamps manually for a frontal AP exam 2 Remove the cap A take the carpus panel B and push it C into the magnetic carpus panel slot D When the carpus panel is connected to the unit the acq...

Page 48: ...ate parameters Click and select Memorize Anatomy setting if you want to save the new parameter settings as future default settings Preparing and Positioning the Adult and Pediatric Patient To prepare...

Page 49: ...ve the X ray room and close the door You must keep visual contact with the patient during acquisition IMPORTANT Make sure that the hand wrist and forearm are in a completely vertical position hand is...

Page 50: ...the nasion support and release the patient X ray Dose Emission Information Compliance with EURATOM 97 43 Directive You can right click on each image to display the estimated emitted dose received by...

Page 51: ...rried out by an authorized service technician Controlling the Image Quality To maintain optimum image quality it must be controlled once a month To control the image quality follow these steps 1 On yo...

Page 52: ...48 Chapter 6 Maintenance 3 Select the desired test and follow the on screen instructions...

Page 53: ...n interface reaches zero Thermal security Cooling in progress Wait until the Generator Cooling Indicator on the Acquisition interface reaches zero Release handswitch The acquisition has ended Release...

Page 54: ...50 Chapter 7 Troubleshooting...