Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04
49
7
Troubleshooting
Quick Troubleshooting
Occasionally, malfunctions can occur during use in the event of an incorrect action. An error
message is displayed on the popup on the
Acquisition
interface
System Status Screen
.
The following table lists the information messages, their description and the action to take:
Table 1
Error Code Message
Table 2
Information messages
Error Code
Error Message
Description
Action
Err_S_GEN_36865
Handswitch was
released before the
end of exposure.
The user has released the
exposure button
prematurely.
Relaunch the acquisition and hold
the exposure button until the end of
the acquisition cycle.
Message
Description
Action
X-ray tube cooling
Cooling in progress.
Wait until the
Generator Cooling
Indicator
on
the
Acquisition
interface reaches zero.
Thermal security
Cooling in progress.
Wait until the
Generator Cooling
Indicator
on
the
Acquisition
interface reaches zero.
Release handswitch
The acquisition has ended.
Release the exposure button of the X-ray
remote control.
Start the acquisition
The acquisition has been
launched.
Keep pressing and holding the exposure button.
Upgrade of the firmware
The system is being upgraded.
Wait for completion of the upgrade.
I 9
The Cephalometric sensor is in
the cooling process.
Wait until the
Ready
Indicator
LED
is green.
I 5
The position of the head clamps
do not match the selected exam
program.
Position the head clamps manually for the
selected exam.
I 17
The sensor is in the positioning
process for either panoramic, 3D
or Ceph mode.
Wait for the end of the positioning process.
If an Error Code message is displayed, the malfunction persists
or more serious conditions occur, contact a qualified technician.
When you call the qualified technician have the following
information ready:
Model Serial Number
Error Code Message
Summary of Contents for CS 8100 Access
Page 6: ...2 Chapter 1 Conventions in This Guide...
Page 26: ...22 Chapter 4 Getting Started...
Page 52: ...48 Chapter 6 Maintenance 3 Select the desired test and follow the on screen instructions...
Page 54: ...50 Chapter 7 Troubleshooting...
Page 56: ...SM987 Ed04 2019 02 For more information visit www carestreamdental com...