CARESONO
PadScan HD 5 User’s Manual
26
screen after the calibrations succeed.
9.
Tap
to exit.
NOTE:
You may take a normal scan on the phantom in the calibration mode to verify the calibration
result. (Compare the scan volume with the phantom labeling volume.)If the volume values
deviate greatly from the acceptable calibration range, perform another calibration.
The calibration can only be done on Dansk Tissue Equivalent Phantom.
The Dansk Phantom should be equipped with the probe holder made by Caresono.
The probe head should be immersed by the water on the phantom surface completely; the
probe head should touch the phantom surface.
The calibration should be done when the main unit is powered on at least 40 minutes.
The calibration should be done when the AC power is on. (Shouldn’t on battery only.)