Page 36 of 52
RD-OI-48 A/0
Table 11 Inspection item
Item
Period
Method
Description
Remedy
Cables
Daily or
Weekly
Visual
inspection
Check that cables are
not damaged and cover
of cables is not torn.
Contact your distributor or
CareRay support if there are
any problems.
Connectors
Daily or
Weekly
Visual
inspection
Check that the plugs
and locks of connectors
are not loose.
Fully insert the cables and
lock them.
Detector
Daily or
Weekly
Visual
inspection
Check that the cover
and parts are not
damaged and not loose.
Contact your distributor or
CareRay support if there are
any problems.
Test exposure
Daily or
Weekly
Test
exposure
Check the exposure
condition of the product
Contact your distributor or
CareRay support if there are
any problems.
Performance
Monthly
or Yearly
Exposure
a phantom
object or a
resolution
chart
Check the general
performance of the
product
Contact your distributor or
CareRay support if there are
any problems.
We recommend the user uses a phantom with the following information (or equivalent):
Manufacturer: Supertech., Inc.
Item number: 07-646
Dimensions: 17" L
×
14" W
×
2.5" Thick (43cm
×
36cm
×
6 cm)
7.2
Maintenance
If any parts need to be replaced, check the product specifications to make sure it is consistent with the
specifications stated in the operation manual. For system performance and safety purposes, choose the
OEM accessories.
Replace Fuse
Two fuses are installed inside of the power box. In order to prevent electrical accidents (such as an
overcurrent occurring in the AC input), stop using the power box immediately if the fuses break.
Table 12 Fuse specification
Item
Description
Model
023502.5
Manufacturer
Littelfuse Inc.
Amp rating
2.5A
Voltage rating
250V