Chapter 7: Appendices
Bird Sentry™ Blender
20
L1383 Rev. D
Mixed gas stability*
±1% oxygen
Ambient operating conditions
Temperature range
59 to 104°F (15 to 40°C)
Relative humidity range
5-95%, non-condensing
Ambient storage conditions
Temperature range
5 to 122°F (-15°C to 50°C)
Relative humidity range
5-95%, non-condensing
*The delivered oxygen concentration will remain constant within ±1% of the set point value with
constant inlet pressures. The displayed value may vary more than this due to sensor accuracy, age,
of the sensor, environmental conditions and length of time since last sensor calibration.
Alarm Specifications
Pressure differential alarm activation
When supply pressures differ by 20 PSI (1.3 BAR) or more
activation
High low concentration alarm range
18 to 99% oxygen
O
2
Sensors
Oxygen measurement accuracy
±2% oxygen
Linearity
±1% oxygen when temperature and pressure are held
constant
Potential error due to changes in
sensor temperature
±3% oxygen, maximum
Response time at 77°F (25°C) to 90%
of final reading
≤20 seconds
Ambient operating temp. range
59 to 104°F (15 to 40°C)
Ambient operating humidity range
0 to 100% (non-condensing)
Ambient storage temp. range
50 to 122°F (–15 to 50°C)
Monitored gas temp. range
59 to 104°F (15 to 40°C)
Monitored gas humidity range
0 to 100%(non-condensing)
Caution!
The use of oxygen sensors not recommended by CareFusion Corporation for use with the Bird
Sentry blender may result in performance that does not meet the specifications listed in this
manual. CareFusion Corporation recommends that only those sensors labeled as Bird
equipment be used with the Sentry.
This device has not been tested in the presence of gases other than air or oxygen.