Specifications
Infusion Specifications -
Maximum infusion rate can be set as part of the configuration.
0.1ml/h - 150ml/h
5ml syringes
0.1ml/h - 300ml/h
10ml syringes
0.1ml/h - 600ml/h
20ml syringes
0.1ml/h - 900ml/h
30ml syringes
0.1ml/h - 1200ml/h
50ml syringes
The Volume Infused range is 0.0ml - 9990ml.
Bolus Specifications -
Selected maximum rates are shown below
150ml/h
5ml syringes
300ml/h
10ml syringes
600ml/h
20ml syringes
900ml/h
30ml
syringes
1200ml/h
50ml syringes
The default bolus volume can be set as part of the configuration.
Minimum: 0.1ml;
Maximum 100.0ml
Increments of 0.1ml; default 5.0ml
During BOLUS the pressure limit alarms are temporarily increased to
their maximum level.
Critical Volume -
The bolus which can occur in the event of a single internal fault
condition with a 50 ml syringe is :
Maximum Overinfusion - 0.5ml
Purge Specifications -
The purge rate is limited to the maximum rate for the syringe and can
be set as part of the configuration.
100ml/h - 500ml/h.
The purge volume range is 0.5ml - 5ml.
During PURGE the pressure limit alarms are temporarily increased to
their maximum level.
Keep Vein Open (KVO) Rate -
0.1 ml/h - 2.5ml/h.
End Of Syringe Rate -
Stop, KVO (0.1ml/h to 2.5ml/h), or set rate if lower than KVO.
Near End Of Infusion Alarm -
1min - 15min to end of infusion, or 10% of syringe volume, whichever
is smaller.
End Of Infusion (EOI) Alarm -
0.1% - 5% of syringe volume
Electrical Classification -
Class I product. Continuous Mode Operation, Transportable
Maximum Pumping Pressure Limit -
Highest alarm level 1000mmHg (nominal at L-10)
Occlusion Accuracy (% of full scale)* -
Pressure mmHg
L-0
approx.
50 mmHg
L-3
approx.
300 mmHg
L-5
approx.
500 mmHg
L-10
approx.
1000 mmHg
Temp.
23°C
±18%
±21%
±23%
±28%
* - Using most common 50ml syringes under normal conditions
(95% confidence / 95% of pumps).
Battery Specifications -
Rechargeable sealed NiMH. Automatically charges when the pump is
connected to AC power.
Mean Time To Battery Empty from fully charged @ 5ml/h & 20°C under
normal conditions is 6 hours*
*95% lower confidence interval of 5 hours 50 minutes
Charging takes 2½ hours from discharge to 90% charge.
Memory Retention -
The electronic memory of the pump will be retained for more than 6
months when not powered up.
System Accuracy -
Volumetric Mean +/- 2% (nominal).
Derating -
Tempe/- 0.5% (5 - 40ºC)
High Rates +/-2.0% (rates > syringe volume/h eg.
>50ml/h in a 50ml syringe.)
Important: System accuracy is +/-2% typical by volume as measured
using the trumpet curve test method defined in IEC/EN60601-2-24 at
rates of 1.0ml/h (23ºC) and above when the pump is used with the
recommended syringes. Caution: Infusion volume accuracy may be
compromised at rates below 1.0ml/h. Differences in factors such as
size and plunger force in compatible syringes can cause variations in
accuracy and trumpet curves. See also trumpet curves section in this
manual.
Fuse Type -
2 x T 1.25A, slow blowing.
AC Power Supply -
115 - 230VAC, 50 - 60Hz, 20VA (nominal).
Dimensions -
310 mm (w) x 121 mm (h) x 200 mm (d). Weight: 2.7 kg (excluding power
cable).
Alarm Conditions -
Drive Disengaged
Occlusion
Check Syringe
Battery Low / Battery Empty
Near End Of Infusion
End of Infusion
AC Power Fail
Internal Malfunction
Attention (Nurse Callback)
Titration not confirmed
Environmental Specifications -
Operating Temperature
+5°C - +40°C
Operating Relative Humidity
20% - 90%
Operating Atmospheric Pressure
700hPa - 1060hPa
Transport & Storage Temperature
-30°C - +50°C
Transport & Storage Relative Humidity
10% - 95%
Transport & Storage Atmospheric Pressure
500hPa - 1060hPa
Electrical/Mechanical Safety -
Complies with IEC/EN60601-1 and IEC/EN60601-2-24.
Protection against fluid ingress -
IPX1 - Protected against vertically falling drops of water.
EMC -
Complies with IEC/EN60601-1-2
and IEC/EN60601-2-24.
Conversion Factor
Dosing Conversion:
1.0 µg = 1000 ng
1.0 mg/h = 24.0 mg/24 h
1.0 mg/min = 60.0 mg/h
1.0 mg = 1000 µg
Volume / Unit Time = Dose Rate / Concentration
1.0 ml/h = 1.0 mg/h / 1.0 mg/ml
The formula is:
Volume/Rate =
(.......µg/kg/min) x (.......kg) x (60 min/h) = µg/h = ml/h
conc. in mg/ml x 1000 µg/mg µg/ml
Drug Units Available:
ml/h, ng/min, ng/kg/min
µg/min, µg/kg/min, µg/h, µg/kg/h, µg/24h, µg/kg/24h
mg/min, mg/kg/min, mg/h, mg/kg/h, mg/24h, mg/kg/24h,
g/h, g/24h, U/min, U/kg/min, U/h, U/kg/h, U/24h, U/kg/24h,
kU/24h, mmol/h