![CareFusion Alaris TIVA User Manual Download Page 22](http://html.mh-extra.com/html/carefusion/alaris-tiva/alaris-tiva_user-manual_2557176022.webp)
1000DF00330 Issue 4
21/34
Alaris® TIVA Syringe Pump Configured Options Record
General Options
Enter the pump-specific information for your records on a copy of this page.
Syringes Enabled
Make
Size(s)
Make
Size(s)
Hospital Name
Serial No.
Software Version
Approved by
Configured by
Date
Date
Option
Default
Range
Setting
Software Version
1.6.2 & 2.1.0
1.9.x &
2.3.x and above
NURSE CALL FITTED
Disabled
Disabled
Enabled/Disabled
NURSE CALL INVERT
Disabled
Disabled
Enabled/Disabled
RS232 SELECTED
Disabled
Disabled
Enabled/Disabled
NEOI WARNING
1min
5mins
1min - 15mins
EOI POINT
1.0%
1.0%
0.1% - 5% of syringe volume
KVO AT EOI
Enabled
Enabled
Enabled/Disabled
KVO RATE
1.0ml/h
1.0ml/h
0.1ml/h - 2.5ml/h
BACK OFF
Disabled
Enabled
Enabled/Disabled
AC FAIL
Enabled
Enabled
Enabled/Disabled
PRESSURE DISPLAY
Disabled
Enabled
Enabled/Disabled
PRESSURE DEFAULT
L-5
L-3
L0 - 10(50mmHg -1000mmHg)
WEIGHT
70.0Kg
70.0Kg
0.01Kg - 250Kg
PURGE RATE
200ml/h
200ml/h
100ml/h - 500ml/h
PURGE VOLUME LIMIT
2.0ml
2.0ml
0.5ml - 5.0ml
PURGE SYRINGE
Enabled
Disabled
Enabled/Disabled
HANDS FREE BOLUS
Enabled
Enabled
Enabled/Disabled
DEFAULT BOLUS VOL
5.0ml
5.0ml
0.1ml - 100ml
DEFAULT BOLUS RATE
1200ml/h
1200ml/h
150ml/h - 1200ml/h
MANUAL BOLUS
Disabled
Enabled/Disabled
CALLBACK TIME
2.0mins
0.1mins - 15.0mins
EVENT LOG DISPLAY
Disabled
Enabled
Enabled/Disabled
BATTERY ICON
Enabled
Enabled/Disabled
AUDIO VOLUME
Medium
Medium
Low, Medium, High
AUTO NIGHT MODE
Enabled
Enabled
Enabled/Disabled