Replacement of the Complete Housing Assembly, Force Sensor Assembly or Logic Board Assembly in a
Syringe Module with version
7
or earlier software must be performed at the depot. Refer to the note in
section
7.4
for further information.
• The Top disk Holder, Internal Frame, Shaft Support and various Syringe Barrel Clamp Assembly parts
identified here are not backwards compatible with a Syringe Barrel Clamp Assembly having a Park
chapter for part identification and replacement instructions.
• Newer versions of the Park Position Collar and Barrel Clamp are not interchangeable with earlier
versions. If either part requires replacement and the hex head bit,
1
/
16
HX
,
1
15
/
16
” length tool was used to
remove the Park Position Collar
0.062
” setscrew, both parts must be replaced.
Order the Front Case “Kit” when replacing an “old version” Front Case, Syringe Barrel Clamp Assembly (or
associated part), or Top Disk Holder (reference Figure
Order the Front Case “Assembly” when replacing a “new version” Front Case, Syringe Barrel Clamp Assembly
(or associated part), or Top Disk Holder (reference Figure
The Base Disk Holder can be relaced as an individual part only if it is used with a “new version” Front Case. If
an “old version” Front Case is in use, order item 9, Front Case “Kit” (reference Figure
Not sold by Car
efusion
.
CAUTIONS
ILLUSTRATED PARTS BREAKDOWN
Syringe Module
7A-6
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
Table 7A-1.
Parts List - Syringe Module
(Continued)
Item
Part Number
Description
QTY
496
499
645
646
2003
2004
2006
321035
148013-000
125569
147077-100
147078-100
147080-100
10010720
10010199
10012555
10012398
148089-001
Screw, 4-40 x
3
/
4
, PNH PHH
Silicone Grease, Dow Corning Molykote 33, or equivalent (use with
item
410
)
Label, Patent
Label, Serial Number Replacement
IUI Connector Kit, Right
(Includes: Right Gasket, Right
IUI
Connector)
IUI Connector Kit, Left
(Includes: Left Seal, Left
IUI
Connector)
Latch Kit
(Includes: Latch, Compression Spring, Leaf Spring, Support)
Directions for Use:
PCA Module (V7)
Syringe Module (V7)
Alaris
®
System (supports Guardrails
®
Suite MX):
Electronic Copy
Printed Copy
Set Compatibility Card
6
As required
1
1
1
1
1
1
1
Summary of Contents for Alaris 8110 series
Page 9: ...1 GENERAL INFORMATION...
Page 10: ......
Page 15: ...2 CHECKOUT CONFIGURATION...
Page 16: ......
Page 27: ...3 PREVENTIVE MAINTENANCE...
Page 28: ......
Page 31: ...4 PRINCIPLES OF OPERATION...
Page 32: ......
Page 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Page 42: ......
Page 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Page 72: ......
Page 105: ...6 TROUBLESHOOTING...
Page 106: ......
Page 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Page 116: ......
Page 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Page 144: ......
Page 171: ......