Warning! 

To minimize the risk of fire, electric shocks or bodily injury: 

 

1.

 

The product must only be used for its intended purpose as described in this manual. 

The product may not be combined, assembled or repaired with other parts (e.g. pump 

and mattress), accessories or spare parts other than those described in this manual or 

other documentation from Care of Sweden. 

 

2.

 

Do not use the product in wet rooms or where there is a risk of the pump coming into 

contact with water or another liquid. Never handle a pump that has come into contact 

with water/liquid. Pull the plug out of the electrical socket immediately and send the 

product to an authorized service technician for servicing.  

 

3.

 

DO NOT USE in the vicinity of or in contact with fire sources/hot surfaces, such as fire, 

burning cigarettes or hot lamps. Although the product has undergone fire testing, it 

could become damaged if it comes into contact with fire sources.  

 

4.

 

Position and use the product in a manner that prevents it from becoming trapped or 

damaged. NOTE: Be particularly aware of trapping damage when using bed rails.  

 

5.

 

Never use the product if the pump power cord or plug is defective, the pump casing is 

damaged or it is not functioning properly. Contact a service technician for examination 

and repair.  

 

6.

 

Regularly check product functionality through a hand check (see section 4.3). Also 

check that the comfort setting is correct as indicated in the table on the pump.  

 

7.

 

Continual monitoring is required when the product is used for individuals needing 

special supervision, such as children.  

 

8.

 

The mattress are protected with a hygiene cover; please try to avoid the simultaneous 

use of multiple hygiene cover with this mattress. Because of the risk that the vapour 

permeability can be destroyed. 

 

9.

 

The hygiene cover does not allow liquid or air to penetrate, but is vapour permeable. 

Make sure that the patient is positioned correctly to avoid the risk of suffocation.  

 

10.

 

Prevent damage to the hygiene cover from sharp objects. 

 

11.

 

If the cover are supplied with handles on the long sides, theses handles are only to be 

used for lifting and moving the mattress itself. All other use will be at your own 

responsibility and is not covered by the product warranty. 

 

12.

 

In order to minimize the risk of wounds occurring on the feet, make sure that the user 

doesn’t come into contact with the hangers of the pump. 

 

IMPORTANT SAFETY PRECAUTIONS 

Read all instructions prior to use 

 

 

 

Summary of Contents for CuroCell 4

Page 1: ...User manual 2019 04 26 CuroCell 4 Alternating pressure system Item number user manual 95 001044...

Page 2: ...roduction 4 Product description 6 Assembly Installation 8 Operation 9 Cleaning and reconditioning 11 Storage 12 Maintenance 13 Troubleshooting 14 Technical description 15 Spare parts 17 Options and ac...

Page 3: ...pump power cord or plug is defective the pump casing is damaged or it is not functioning properly Contact a service technician for examination and repair 6 Regularly check product functionality throug...

Page 4: ...ing them immediately to us here at Care of Sweden The system has been tested and approved according to the following standards IEC 600601 1 IEC 60601 1 2 SS EN ISO 12182 SS EN ISO 14971 SS EN ISO 1099...

Page 5: ...u can order a copy from our Customer services by sending your details to info careofsweden se NOTE For certain patients such as amputees are the recommended minimum weight not valid These users may ho...

Page 6: ...on 2 1 Pump and mattress 1 CPR rapid deflation a Head b Foot 2 Pump 3 Mattress 4 Pump connector 1a Tubing system with CPR in head 1b Tubing system with CPR in foot With the optional heel section is th...

Page 7: ...t care repositioning transferring in and out of bed Turn off this function by pressing the button It will return automatically to cycling after approx 30 minutes 5 Low pressure LOW PRESSURE The pump h...

Page 8: ...d correctly 2 Hang the pump at the foot end of the bed or place it on a flat steady surface Route the power cord so that there is no risk of it causing someone to trip being run over by the bed wheels...

Page 9: ...prescriber or other authorized personnel for appropriate setting 1 Use the comfort dial to set a suitable air pressure based on the settings table on the pump 2 When the green indicator lamp for norm...

Page 10: ...til the pressure is corrected and reset See section 8 for troubleshooting 4 5 CPR In the event of an emergency requiring CPR cardiopulmonary resuscitation Quickly pull the CPR strap remove the quick c...

Page 11: ...alcohol with or without tensides Isopropanol 70 oxidating solutions or chlorine max 1 If using a different agent choose one that does not affect the pump housing MATTRESS COVER Covers can be wiped of...

Page 12: ...tubes and the CPR valve with a mild cleaning solution for this purpose such as alcohol with or without tensides Isopropanol 70 oxidating solutions or chlorine max 1 Let it dry 5 Wipe off and disinfect...

Page 13: ...parts 7 2 Changing fuses 1 Pull the plug out of the electrical socket before replacing fuses 2 Remove the cover for the fuses with a screwdriver Pull out the fuse holder 3 Fit a new fuse of the same t...

Page 14: ...ew minutes Do a hand check Refer to section 4 3 Check that the CPR valve valves are properly closed Mattress moves around Check that the air mattress is fastened to the bed frame with the straps under...

Page 15: ...35 C Storage 15 C 50 C Shipping 15 C 70 C Humidity Operation 20 80 non condensing Storage 10 90 non condensing Atmospheric pressure 700hPa 1060hPa Classification Class II Type BF IP21 MATTRESS SPECIFI...

Page 16: ...ycling Read the user manual Machine wash at 70 C CE marked acc to MDD 93 42 EEC Machine wash at 95 C Directly on bed base Tumble dry Do not iron Recommended user weight Patient information category Wi...

Page 17: ...10 CC 916216 CC 907430 CC 905510 120x200 cm CC 906920 CC 916218 CC 907440 CC 905521 Art no Name Size CC 901310 Cell PU 80x21 cm CC 901311 Cell PU 85x21 cm CC 901312 Cell PU 90x21 cm CC 901313 Cell PU...

Page 18: ...els of the bed NOTE This option is only available with new orders Sheet holder Specialised clips to which the sheet can be attached Six required per mattress 11 2 Accessories Art no Description and si...

Page 19: ...t be submitted to Care of Sweden before the product is returned Please contact Care of Sweden or your local dealer if you have questions regarding warranty and service 12 3 Life expectance This produc...

Page 20: ...ternet www careofsweden com Address Care of Sweden AB P O Box 146 SE 514 23 Tranemo SWEDEN Visit address Fabriksgatan 5A SE 514 33 Tranemo SWEDEN Cargo address Byns v g 4A SE 514 33 Tranemo SWEDEN Sup...

Reviews: