CardioMed Treadmills Operators Manual V 2.0
9
Treadmill Operation
1. Plug in the treadmill to the power source.
2. While both the treadmill and control device are off, connect the treadmill to the control
device via USB or Serial Cable.
3. Once connected, turn on the power to the treadmill first then the control device.
4. Use the control device to select the proper COM Port for operation.
5. If the treadmill losses connection or the COM Port cannot be found, repeat steps 1-4.
Treadmill Placement
The product should be positioned in a place that is stable and horizontal.
1. Check if there is any obstacle in front or behind the front wheels
2. Check that the product is stable and secure, if not tighten screws and adjust levellers.
Operating Environment
The product is designed to operate under the following conditions:
Temperature from +15
o
C until +30
o
C
The instrument was not designed to operate outdoors or in places with high humidity, or where
there is danger of fire or explosion.
The position of the user is in front of the console.
There should exist no environmental conditions (for instance powerful noise, bright light etc)
that would contribute to the natural or intellectual difficulty of the user.
The product should be positioned at an area with enough peripheral space, so that access for
use or maintenance is allowed. It should not be placed near any source of heat and it should
be continuously ventilated.
Alignment of Walking Belt
The alignment of the belt is correct when it moves to the middle of walking surface without
slipping. The alignment can be achieved from the bolt that is on the left hand side of the frame.
In order to move the belt left you turn the screw counter clockwise (quarter turns are
recommended). If the belt is traveling too much to the left then you would turn the screw
clockwise see picture.
Use only the left screw
for alignment
, using both screws to adjust
the tracking can create excessive tension on the belt which can result in decreasing the life
expectancy of the drive motor and walking belt.
Direction of rotating the screw