8
CALIBRATION, Cont
BREAKPOINT
Press
NET
to show the current break point setting.
Press ON/ZERO
to select
break point setting (1 through 99). Press the
NET
key to save setting.
MOTION RANGE
Press the
NET
key to show the current motion range in divisions. Then press the
ON/ZERO
key to select the number of digits to change before motion:
0
through
9
. Press the
NET
key to save.
SAMPLE RATE
Press the
NET
key to show the current sample rate setting. Values of
1
through
12
are valid for the sample rate. Press the
ON/ZERO
key to select
1
through
9
samples per second, then press the
NET
key to save. If a sample rate of
10
through
12
is desired, first press the
ON/ZERO
key to select
0
through
2
, then
press the
TARE
key to advance to the next digit location and press the
ON/ZERO
key to select
10
through
12
samples per second. Press the
NET
key to save.
AUTOMATIC SHUTOFF
Press the
NET
key to show the current setting. Then press the
ON/ZERO
key to
disable the feature or to select the number of minutes of inactivity before turning off
the scale.
0=
disable,
1
to
9=
minutes of inactivity before shutoff. Press the
NET
key to save it.
SLEEP MODE
Press the
NET
key to show the current setting. Then press the
ON/ZERO
key to
disable the feature or to select the number of minutes at zero before going to sleep.
0=
disable,
1
to
9=
minutes of inactivity before sleeping. Press the
NET
key to save
it.
PUSH BUTTON TARE
Press the
NET
key to show the current setting. Then press the
ON/ZERO
key to
toggle between yes and no.
YES=
enable,
no=
disable. Press the
NET
key to save.
The display will display the software revision number and then the gross weight. The setup and
calibration process has been completed and the scale is ready to weigh. Place the test weight back
on the scale to verify calibration.
UnS=
b=
Sr=
ASh=
SLP=
Pbt=