8200-M503-01 Rev-B
•
2XX-IP
5
INSTALLATION, Cont.
Ethernet Cable Installation, Cont.
9. The shield wire is secured when tightening the gland connector nut. See figure below.
10. Do not over-tighten the connector but make certain it is snug.
11. Consult your network administrator for the proper procedure to terminate the cable.
12. After the cable has been terminated, plug the cable into the Ethernet port on the 2XX-IP.
Re-Installing the 205/210 and 225 Rear Panel
After all terminations have been made;
1. Remove the excess cable from the instrument enclosure and securely tighten each of the
cable gland connectors.
•
Do not over-tighten these connectors but make certain they are snug.
•
DO NOT USE TOOLS!
Finger-tighten only!
•
Insure any unused gland connectors are plugged.
2. Make certain no cables or wires are exposed between the main housing and rear panel
3. Place the rear panel onto the main housing.
4. Secure with the acorn nuts removed earlier.
5. Follow a diagonal pattern when tightening the screws.
Re-Installing the 212G/212GX Front panel
After all terminations have been made;
1. Remove the excess cable from the instrument enclosure and securely tighten each of the
cable gland connectors.
•
Do not over-tighten these connectors but make certain they are snug.
•
DO NOT USE TOOLS!
Finger-tighten only!
•
Insure any unused gland connectors are plugged.
2. Make certain no cables or wires are exposed between the main housing and front panel
3. Place the front panel onto the main housing.
4. Secure with the mounting screws removed earlier.
5. Follow a diagonal pattern when tightening the screws.