When the Kendall SCD™ 700 series controller detects a fault condition, it:
• Interrupts the normal operation of the controller
• Deactivates all valves to vent the air from the garment
• Displays a fault code
• Sounds an audible error
Once you confirm the cause of the error and remedy the problem,
you can reset the error.
Some errors will remain active until you turn off the controller or the
battery runs out of charge (if operating on battery power). If a Garment Mismatch error is
triggered, the user may remedy the problem by pressing the corresponding A and B Button(s).
The Kendall SCD™ 700 series controller has four error types:
User Resettable, Manual Reset Required, Service Required, and Recharge Battery.
User Resettable
This type of error allows the user to remedy the issue and resume operation by pressing the A and B button(s) corresponding with the port
affected without powering the unit down. For this type of error, a check mark will be shown indicating which port is the area of concern. A
yellow triangle indicates a low concern error. If the triangle is red, it indicates an error related to abnormally high pressure. If the error persists,
then the controller requires service.
Manual Reset Required
This error can be corrected by the user but requires the device to be powered off and on. If the error persists, then the controller
requires service.
Service Required
This error code occurs because of a failed internal component. It cannot be addressed by the user.
Recharge Battery
This error can be corrected by the user plugging in the controller into an AC power outlet.
Error key for
Kendall SCD™ 700
series controller
Summary of Contents for Kendall SCD 700
Page 5: ...GET SMART ABOUT COMPRESSION SMART COMPRESSION THERAPY Kendall SCD 700 SMART COMPRESSION...
Page 7: ...PROTECTS ADAPTS DETECTS SUSTAINS EDUCATES business card cardinalhealth com getsmartcompression...
Page 9: ...Kendall SCD 700...
Page 11: ...2 Kendall SCD 700 1 2 3 4 Kendall SCD 700 1 a b c d 2 3 4 5 1 2 3...
Page 12: ...3 1 Kendall SCD 700 2 3 4 5 6 7 8 9 10 CE...
Page 14: ...5 1 2 3 6 7 4 5 1 5 1 3 2 6 3 A 7 B 4 AC...
Page 15: ...6 I 1 2 B A AC AC II...
Page 16: ...7 LED LED 2 A B A B A B A B A B 1 1 A B...
Page 17: ...8 A B A B A B 1 1 A B A B 2 2 A B...
Page 19: ...10 3...
Page 21: ...12 II Kendall SCD 700 1 I 8 10 12 24 999 40 0 4 B A...
Page 22: ...13 3 00 00 Kendall SCD 700 5 6 5 6...
Page 23: ...14 7 8 3 7 8...
Page 24: ...15 4 30 8 10 12 24 1 A 1 8 8 24 24 30 1 B 9 2 A 3 B X 4 A B 5 0 6...
Page 25: ...16 9...
Page 26: ...17 8 2 5 1 3 0 5 1 10 2 2...
Page 28: ...19 AC 1 2 3 0 100 3 15 40 2 1 3 15 11 11 AC 4 8 6 8 50 25 77 F...
Page 29: ...20 Kendall SCD 700 Li 20 60 4 F 140 F IV A B E5...
Page 30: ...21 A B E1 A B II...
Page 31: ...22 A B A B E1 12 0kPa 90 mmHg 2 4 0 k P a 180 mmHg A B E2 10 6 3 kPa 47 mmHg 5 8 7 kPa 65 mmHg...
Page 33: ...24 E5 E6 E7 A B E8 2 7 kPa 20 mmHg...
Page 34: ...25 E9 E9 5 41 F 55 131 F E10 A B E12 10 1 3 kPa 10 mmHg...
Page 35: ...26 E13 0 7 kPa 5 mmHg 15 AC...
Page 36: ...27 V Kendall SCD 700 Kendall SCD 700 12 Covidien Kendall SCD 700...
Page 38: ...29 VI ESD 1 2 3 5 Torx T15 4 5 6 12 1 2 1 Torx T15 2 3 4...
Page 39: ...30...
Page 46: ...37 12 19 20 16 1 17 18...
Page 48: ...39 14...
Page 49: ...40 15 CPU...
Page 51: ...1...