Section 2 – Instructions for Customer Returned Units (continued)
PREPARATION (continued)
TUBING SET LEAK TEST (continued):
Turn ball valve to open position, so that the knob position is in-line with valve
body (see picture below).
Adjust the input air pressure so that a pressure of between 125mmHg (2.4 psi)
and 135mmHg (2.6 psi) is indicated on the fixture’s pressure gauge, and allow to
hold for approximately 2 seconds.
Close ball valve, so that the knob position is perpendicular to valve body (see
picture below). Activate the stopwatch.
Wait 15 seconds and observe the pressure gauge. Because the test was
conducted with known good tubing set, the pressure should not fall below
89mmHg (1.72 psi). If pressure falls below 89mmHg, contact your sales rep for
replacement tubing set Leak Test Fixture.
Summary of Contents for Kendall SCD 700
Page 5: ...GET SMART ABOUT COMPRESSION SMART COMPRESSION THERAPY Kendall SCD 700 SMART COMPRESSION...
Page 7: ...PROTECTS ADAPTS DETECTS SUSTAINS EDUCATES business card cardinalhealth com getsmartcompression...
Page 9: ...Kendall SCD 700...
Page 11: ...2 Kendall SCD 700 1 2 3 4 Kendall SCD 700 1 a b c d 2 3 4 5 1 2 3...
Page 12: ...3 1 Kendall SCD 700 2 3 4 5 6 7 8 9 10 CE...
Page 14: ...5 1 2 3 6 7 4 5 1 5 1 3 2 6 3 A 7 B 4 AC...
Page 15: ...6 I 1 2 B A AC AC II...
Page 16: ...7 LED LED 2 A B A B A B A B A B 1 1 A B...
Page 17: ...8 A B A B A B 1 1 A B A B 2 2 A B...
Page 19: ...10 3...
Page 21: ...12 II Kendall SCD 700 1 I 8 10 12 24 999 40 0 4 B A...
Page 22: ...13 3 00 00 Kendall SCD 700 5 6 5 6...
Page 23: ...14 7 8 3 7 8...
Page 24: ...15 4 30 8 10 12 24 1 A 1 8 8 24 24 30 1 B 9 2 A 3 B X 4 A B 5 0 6...
Page 25: ...16 9...
Page 26: ...17 8 2 5 1 3 0 5 1 10 2 2...
Page 28: ...19 AC 1 2 3 0 100 3 15 40 2 1 3 15 11 11 AC 4 8 6 8 50 25 77 F...
Page 29: ...20 Kendall SCD 700 Li 20 60 4 F 140 F IV A B E5...
Page 30: ...21 A B E1 A B II...
Page 31: ...22 A B A B E1 12 0kPa 90 mmHg 2 4 0 k P a 180 mmHg A B E2 10 6 3 kPa 47 mmHg 5 8 7 kPa 65 mmHg...
Page 33: ...24 E5 E6 E7 A B E8 2 7 kPa 20 mmHg...
Page 34: ...25 E9 E9 5 41 F 55 131 F E10 A B E12 10 1 3 kPa 10 mmHg...
Page 35: ...26 E13 0 7 kPa 5 mmHg 15 AC...
Page 36: ...27 V Kendall SCD 700 Kendall SCD 700 12 Covidien Kendall SCD 700...
Page 38: ...29 VI ESD 1 2 3 5 Torx T15 4 5 6 12 1 2 1 Torx T15 2 3 4...
Page 39: ...30...
Page 46: ...37 12 19 20 16 1 17 18...
Page 48: ...39 14...
Page 49: ...40 15 CPU...
Page 51: ...1...