Cardinal Health Kendall SCD 700 Manual Download Page 1

CUSTOMIZED 

THERAPY.

CLINICALLY 

PROVEN.

Introducing 

Kendall SCD

 700 Sequential 

Compression System Featuring 

Patient Detection

Summary of Contents for Kendall SCD 700

Page 1: ...CUSTOMIZED THERAPY CLINICALLY PROVEN Introducing Kendall SCD 700 Sequential Compression System Featuring Patient Detection...

Page 2: ...s tracking of patient therapy by detecting that the garments are actively worn Animated alerts Icons show the cause of an error and how it can be resolved Durability Limited liquid ingress rating of I...

Page 3: ...le is applied to the patient s leg 2 The sleeve deflates holding 6 mmHg of pressure in the middle chamber of the calf 3 Pressure is held and monitored for up to 60 seconds to allow evaluation of calf...

Page 4: ...ormity of the leg Suspected pre existing deep venous thrombosis The system may not be recommended for use with foot cuffs on patients with Conditions where an increase of fluid to the heart may be det...

Page 5: ...GET SMART ABOUT COMPRESSION SMART COMPRESSION THERAPY Kendall SCD 700 SMART COMPRESSION...

Page 6: ...M WITH HOSPITAL ASSOCIATED VTE 60 THE PROBLEM WITH CONVENTIONAL COMPRESSION Ability to customize cycles for the patient2 Blood movement2 Nurse time3 Compliance4 1 Jha AK Larizgoitia I Audera Lopez C P...

Page 7: ...PROTECTS ADAPTS DETECTS SUSTAINS EDUCATES business card cardinalhealth com getsmartcompression...

Page 8: ...ardinal Health LOGO Kendall SCD Smart Compression and Patient Sensing are trademarks or registered trade marks of Cardinal Health and may be registered in the US and or in other countries All other ma...

Page 9: ...Kendall SCD 700...

Page 10: ...1 2 2 2 2 3 3 5 I 6 6 7 7 9 9 10 11 II 12 12 13 15 15 16 III 18 19 20 IV 20 V 27 27 27 28 28 28 28 VI 29 VII 31 VIII 36...

Page 11: ...2 Kendall SCD 700 1 2 3 4 Kendall SCD 700 1 a b c d 2 3 4 5 1 2 3...

Page 12: ...3 1 Kendall SCD 700 2 3 4 5 6 7 8 9 10 CE...

Page 13: ...4 BF 15 85 20 55 WEEE 12 5 60 60 UL60601 1 ANSI AAMI ES60601 1 2005 CAN CSA C22 2 NO 601 1 CAN CSA C22 2 NO 60601 1 2008...

Page 14: ...5 1 2 3 6 7 4 5 1 5 1 3 2 6 3 A 7 B 4 AC...

Page 15: ...6 I 1 2 B A AC AC II...

Page 16: ...7 LED LED 2 A B A B A B A B A B 1 1 A B...

Page 17: ...8 A B A B A B 1 1 A B A B 2 2 A B...

Page 18: ...9 E12 IV A B 6 0 kPa 45 mmHg 17 3 kPa 130 mmHg Kendall SCD 700 Covidien Vascular Refill Detection 30 3 20 60 2...

Page 19: ...10 3...

Page 20: ...11 Kendall SCD 700 74010 XS 9545 S 74011 S 9530 M 74012 M 9736 M 74013 L 9780 L 74021 S 9529 M 74022 M 9789 L 74023 L 74041 S 9545T S 74042 M 9530T M 74043 L 9780T L 5897 S M 5898 L 9528 9595...

Page 21: ...12 II Kendall SCD 700 1 I 8 10 12 24 999 40 0 4 B A...

Page 22: ...13 3 00 00 Kendall SCD 700 5 6 5 6...

Page 23: ...14 7 8 3 7 8...

Page 24: ...15 4 30 8 10 12 24 1 A 1 8 8 24 24 30 1 B 9 2 A 3 B X 4 A B 5 0 6...

Page 25: ...16 9...

Page 26: ...17 8 2 5 1 3 0 5 1 10 2 2...

Page 27: ...18 III AC DC LED3 10 10 1 3 2 AC 1 2 3 100 67 99 34 66 0 33 AC 3 1 2 3 67 100 34 66 34 15 40 3 15 3...

Page 28: ...19 AC 1 2 3 0 100 3 15 40 2 1 3 15 11 11 AC 4 8 6 8 50 25 77 F...

Page 29: ...20 Kendall SCD 700 Li 20 60 4 F 140 F IV A B E5...

Page 30: ...21 A B E1 A B II...

Page 31: ...22 A B A B E1 12 0kPa 90 mmHg 2 4 0 k P a 180 mmHg A B E2 10 6 3 kPa 47 mmHg 5 8 7 kPa 65 mmHg...

Page 32: ...23 A B E2 10 18 0 kPa 135 mmHg 5 21 3 kPa 160 mmHg 10 5 7 kPa 43 mmHg 10 16 7 kPa 125 mmHg 12 4 7 7 3 kPa 35 55 mmHg 12 14 7 20 0 kPa 110 150 mmHg...

Page 33: ...24 E5 E6 E7 A B E8 2 7 kPa 20 mmHg...

Page 34: ...25 E9 E9 5 41 F 55 131 F E10 A B E12 10 1 3 kPa 10 mmHg...

Page 35: ...26 E13 0 7 kPa 5 mmHg 15 AC...

Page 36: ...27 V Kendall SCD 700 Kendall SCD 700 12 Covidien Kendall SCD 700...

Page 37: ...28 1 Kendall SCD 700 0 5 Dispatch 70 0 37 o Precise 0 15 0 15 Spray Nine 7 35 0 23 Sporgon 3 4 Cidex Manu klenz 0 5 Dispatch 70 7 35 0 23 Sporgon Manu klenz...

Page 38: ...29 VI ESD 1 2 3 5 Torx T15 4 5 6 12 1 2 1 Torx T15 2 3 4...

Page 39: ...30...

Page 40: ...EC 60601 1 2 2007 UL E189131 E351453 I BF AP APG IP23 IPX3 EN60529 11 5 6 0 kPa 45 mmHg 17 3 kPa 130 mmHg 13 3 96 m 17 3 cm 6 8 19 6 cm 7 7 11 4 cm 4 5 18 5 cm 7 3 2 3 kg 5 0 100 240 VAC 50 VA 50 60 H...

Page 41: ...32 EMC EMC EMC RF Kendall SCD 700 Kendall SCD 700 RF CISPR 11 1 Kendall SCD 700 RF RF RF CISPR 11 B Kendall SCD 700 IEC 61000 3 2 A IEC 61000 3 3...

Page 42: ...kV 6 kV 8 kV 30 IEC 61000 4 4 2 kV 1 kV 2 kV 1 kV IEC 61000 4 5 1 kV 2 kV 1 kV 2 kV IEC 61000 4 11 0 5 5 UT UT 95 5 40 UT UT 60 25 70 UT UT 30 5 5 UT UT 95 0 5 5 UT UT 95 5 40 UT UT 60 25 70 UT UT 30...

Page 43: ...IEC 61000 4 3 3 Vrms 150 kHz 80 MHz 3 V m 80 MHz 2 5 GHz 3 Vrms 3 V m RF Kendall SCD 700 d 3 5 V1 P d 3 5 E1 P 80 MHz to 800 MHz d 7 E1 P 800 MHz to 2 5 GHz P W d m a RF b 1 80 MHz 800 MHz 2 a AM FM...

Page 44: ...s Kendall SCD 700 RF RF W m 150 kHz 80 MHz d 3 5 V1 P 80 MHz 800 MHz d 3 5 E1 P 800 MHz 2 5 GHz d 7 E1 P 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 17 1 17 2 33 10 3 69 3 69 7 38 100 11 67 11 67 23 33...

Page 45: ...mbly 8 USB connector 9 Muffler 10 Rear Enclosure assembly 11 Power cord see spare parts insert for country specific cord 12 Bed hook 13 Power cord door 14 Bed hook spring 2X 15 Bed hook Pivot cover 16...

Page 46: ...37 12 19 20 16 1 17 18...

Page 47: ...J2 CPU J2 POWER SUPPLY BOARD FAN J1 EQUIPOTENTIAL LUG MAINS VOLTAGE INPUT 100 240 VAC 50 60 Hz J9 J3 POWER SUPPLY J4 AC INLET FUSES J6 J5 J1 TO J4 SUPPLY POWER FLOW DIRECTION CPU TO J6 COMPRESSOR AIR...

Page 48: ...39 14...

Page 49: ...40 15 CPU...

Page 50: ...tered trademarks of Covidien AG Trademark of its respective owner Other brands are trademarks of a Covidien company 2011 Covidien Printed in China Covidien llc 15 Hampshire Street Mansfield MA 02048 U...

Page 51: ...1...

Page 52: ...on the end user s computers and is capable of communicating with Covidien medical devices The Client is the broker of communication between the Covidien device and the DMP receiving data from the Serv...

Page 53: ...tion 16 5 2 Step 2 Enter Serial Number 18 5 3 Step 3 Pre Update Check 19 5 4 Step 4 Software Update 20 5 5 Step 5 Post Update Check 23 5 6 Step 6 Update Confirmation 24 6 User Interface Navigation 25...

Page 54: ...ucted The user performing the application installation on the personal computer must have sufficient rights to install and configure software on the user s personal computer The following materials ar...

Page 55: ...SCD 700 series controller herein after referred to as Kendall SCDTM 700 Updater installation Log into the computer that will serve as the service personal computer Place the CD containing the Kendall...

Page 56: ...ed by th Kendall SCD pdater inst he Installin ll appear I few minutes zard autom ogresses to creen n the Welc evice Mana gent Setup creen selec oceed to th creen come to agement Wizard ct Install t in...

Page 57: ...greement select I agree to proceed to the next screen On the Choose Start Menu Folder screen select Install The Installing screen will appear It may take a few minutes before the Wizard automatically...

Page 58: ...the Kenda 00 Updater he Installin ll appear I few minutes zard autom ogresses to creen asked Do y ow the follo ogram from nknown sou ublisher to m hanges to th omputer Se pleting the agement Wizard c...

Page 59: ...Wizard screen select Next to proceed to the next screen On the Welcome to the Covidien Kendall SCD 700 Updater Setup Wizard screen select Next to proceed to the next screen On the Ready to Install sc...

Page 60: ...close the wizard 3 Launching the Software Updater for Kendall SCD 700 Series Controller Note Training and Login credentials username and password are required prior to launching the Kendall SCDTM 700...

Page 61: ...dien Kendall SCD 700 Updater Software Updater Enter User Name and Password in the login screen then select Next Passwords are case sensitive Note The personal computer should be connected to the inter...

Page 62: ...has an internet connection by navigating to an external website Re enter your User Name and Password If you are still unable to login please contact Covidien technical support After login read the le...

Page 63: ...ftware to the personal computer Note An external internet connection is also required to download the Kendall SCD 700 Updater Software Updater Home After the Kendall SCD 700 Updater has been downloade...

Page 64: ...s listed in Section 5 Or 2 In disconnected mode personal computer not connected to the external internet or centralized server the user may Log Out of the Kendall SCD 700TM Updater and move to a locat...

Page 65: ...connection the personal computer should first be prepared following the steps listed in Section 4 1 To start controller software updates launch the Kendall SCD 700TM Updater and login following the s...

Page 66: ...ct series co Step 1 con Preparation Remove series co are two s removed Use Tor the screw date Prepa ntinued U n t the perso er t the Kenda ontroller to ntinued U n e the Kenda ontroller co screws that...

Page 67: ...t USB cable to a USB port located on the personal computer Step 1 continued Update Preparation The user s country is based on the user s profile and is set during user registration The user s country...

Page 68: ...er so that the serial number label is facing you Enter the serial number into the Enter Serial Number field Hit Enter on the keyboard Re enter the same serial number into the Re Enter Serial Number fi...

Page 69: ...inued Pre Update Check Several seconds after the Controller has been turned on the Kendall SCDTM 700 Updater will notify the user that the Pre update function check has passed Press the On Standby but...

Page 70: ...he Home Screen Step 4 Software Update continued The Select Software Version field will be automatically populated for users who are not Covidien employees On the Kendall SCD 700 series controller simu...

Page 71: ...version WARNING Once the Start Update button has been selected the Kendall SCD 700 series controller SHOULD NOT be disconnected from the personal computer for the remainder of the update If the USB c...

Page 72: ...ss bar will depict the status of the software upgrade Step 4 continued Software Update Once the software upgrade is complete the user will be notified and the Next selection will appear at the bottom...

Page 73: ...continued Post Update Check Several seconds after the Kendall SCD 700 series controller has been turned on the Kendall SCDTM 700 Updater will notify the user that the Post Update function check has p...

Page 74: ...able can now be safely removed from the Kendall SCD 700 series controller Reattach the cord cover to the Kendall SCD 700 series controller User can select o Update Another Device to update another Ken...

Page 75: ...ralized server Red indicates that the personal computer is not connected to the centralized server Controller Connectivity Indicator Green indicates that the personal computer is connected to the Kend...

Page 76: ...g the Help link will guide you to additional support information for the Kendall SCDTM 700 Updater and the Kendall SCD 700 Series Controller Operator and Service Manual The Help Screen will not displa...

Page 77: ...he SCDTM 700 Updater is in Disconnected Installation mode the user will have 48 hours to perform the Kendall SCD 700 series controller software upgrades Time remaining is shown in an hour h and minute...

Page 78: ...ment Agent Setup window Select OK on the Device Management Agent Installation window On the Welcome to the Covidien Kendall SCD 700 Updater Setup Wizard screen select Next to proceed to the next scree...

Page 79: ...indow is displayed Press OK to complete the software installation Restart the personal computer Confirm that the personal computer has an internet connection by navigating to an external website Launc...

Page 80: ...Passwords are case sensitive If you are still unable to login contact Covidien technical support If during login the Please shutdown Kendall SCDTM 700 Updater pop up window is displayed Shutdown the K...

Page 81: ...ng is 0h 0m is displayed It is been more than 48 hours since the last connection to the centralized server Logout of the Kendall SCDTM 700 Updater Close the Kendall SCD 700 Updater Application Confirm...

Page 82: ...Number is incorrect pop up window is displayed Select Okay Please re attempt serial number entry for the attached Kendall SCDTM 700 series controller If the Does not Match is displayed The first seria...

Page 83: ...ler Turn off controller Wait for 5 seconds Select Reattempt Check If the Error Screen continues to display after following the prompts contact Covidien technical support 7 6 Step 4 Troubleshooting Ste...

Page 84: ...to the controller and the computer Check the AC power connection to the controller Restart controller Wait for the controller to enter Garment Detection legs blinking Turn off controller Check Confirm...

Page 85: ...d it will need to be returned for service It is important to reattempt the update at this time Select Reattempt Update To Reattempt Software Update Check the USB connections on controller and computer...

Page 86: ...on to the controller Turn Off controller Wait for 5 seconds Select Reattempt Check If the Error Screen continues to display after following the prompts contact Covidien technical support 7 8 Notice Se...

Page 87: ...orrectly Screen does not display correctly If the Kenall SCDTM 700 Updater screens are not displayed correctly Change the computer font size to 100 by selecting Control Panel Appearance and Personaliz...

Page 88: ...indow is displayed Check the USB connections to the controller and the computer Check the AC power connection to the controller Restart controller Wait for the controller to enter Garment Detection le...

Page 89: ...SCD 700 series controller display screen shows 7 01 Is 7 01 this the current software revision No 7 01 is not the current controller software revision 1 08 01 is the software revision 6 Why can t I ch...

Page 90: ...e in any way iii merge the Software with any other software or product not supplied by COVIDIEN iv use copy sell sublicense lease rent loan assign convey or otherwise transfer the Software except as e...

Page 91: ...d to replacement or refund and shall not encompass any other damages No dealer distributor agent or employee of COVIDIEN is authorized to make any modification or addition to the warranty and remedies...

Page 92: ...r or to Customer If any provision of this Agreement shall be held by a court of competent jurisdiction to be illegal invalid or unenforceable the remaining provisions shall remain in full force and ef...

Page 93: ...table This type of error allows the user to remedy the issue and resume operation by pressing the A and B button s corresponding with the port affected without powering the unit down For this type of...

Page 94: ...an 43 mmHg for 10 consecutive cycles Check for leaks in the sleeve or the tube connections Low Pressure Foot Cuffs User Resettable Foot cuff pressure is less than 125 mmHg after 10 consecutive cycles...

Page 95: ...triggered Return to Cardinal Health for service Compressor Error Service Required If the compressor electrically malfunctions this error will be displayed Service technician only Verify that the compr...

Page 96: ...and ESSENTIAL TO CARE and Kendall SCD are trademarks or registered trademarks of Cardinal Health and may be registered in the US and or in other countries All other marks are the property of their re...

Page 97: ...Instruction Guide for Cleaning Testing and Packaging Kendall SCD 700 Sequential Compression System Controller and Tubing Set...

Page 98: ...Instructions for Customer Returned Units 5 Reference Documents 5 Equipment List 5 Preparation 5 Tubing Set 7 Visual Inspection 7 Cleaning 7 Leak Test 7 Kendall SCD 700 Controller 9 Visual Inspection 9...

Page 99: ...0 962 9888 www covidien com APPLY DME LABEL TO KENDALL SCD 700 CONTROLLER Remove home controller from the packaging and take out of the polybag and AIR PAQ inflatable packaging Create a label that pro...

Page 100: ...me controller in the polybag and AIR PAQ inflatable packaging Place the tubing sets into the original package the way it was received see picture below Place the QuickStart Guide on top of the AIR PAQ...

Page 101: ...x and confirm all labels are visible on the outside of the shipping box as shown below o Serial label on the top of the shipping box o Shipping label on the front wrapping on the left side o Return la...

Page 102: ...n receipt of the fixture to confirm acceptable fixture functionality o Confirm the tubing set Leak Test Fixture s certification from the fixture manufacturer o Confirm pressure gauge is calibrated If...

Page 103: ...2 4 psi and 135mmHg 2 6 psi is indicated on the fixture s pressure gauge and allow to hold for approximately 2 seconds Close ball valve so that the knob position is perpendicular to valve body see pic...

Page 104: ...ubing set CLEANING Clean the customer returned tube set as per the Kendall SCD 700 system Operation and Service Manual All visible extraneous markings or materials should be removed by the cleaning st...

Page 105: ...d on the fixture s pressure gauge and allow to hold for approximately 2 seconds Close ball valve so that the knob position is perpendicular to valve body see picture below Activate the stopwatch Wait...

Page 106: ...ee below Minor surface scratches or scuffs are acceptable If any marks are present that obscure or interfere with the display of the icons or the clear membrane on the control panel send the controlle...

Page 107: ...power to enter test mode See front panel display below Enter Test Access Mode and conduct the general function test test mode T2 as per the Kendall SCD 700 system Operation and Service Manual Verify...

Page 108: ...1 1033373 Covidien 1 800 962 9888 www covidien com Polybag for tubing set 2 F123450 Covidien 1 800 962 9888 www covidien com Twist Ties Blue Plastic 1 F150023 Covidien 1 800 962 9888 www covidien com...

Page 109: ...and use a twist tie to secure o Place the Kendall SCD 700 controller in the AIR PAQ inflatable packaging and close the AIR PAQ inflatable packaging o Place the QuickStart Guide on top of the AIR PAQ...

Page 110: ...ial Number Reason for service return Contact information for Estimate approval if applicable Name Phone Number Email address Repackage the Kendall SCD 700 controller carefully in its original box List...

Page 111: ...Quick Start Guide Kendall SCD 700 Sequential Compression System Leg Sleeve s Leg Sleeve s Controller Foot Cuff s Tubing Foot Cuff s Controller Tubing Set Garments packaged separately H29525...

Page 112: ...nough room for two fingers to fit between sleeve and leg Wrap and secure cuff should fit snug like a sneaker 2011 Covidien Printed in China Covidien llc 15 Hampshire Street Mansfield MA 02048 USA www...

Page 113: ...do Sistema de compresi n secuencial Kendall SCD 700 Botas de compresi n Botas de compresi n Controlador Manguitos de pie Mangueras Manguitos de pie Controlador y juego de mangueras Prendas empacadas p...

Page 114: ...na Una vez envuelto y asegurado el manguito debe quedar ajustado como el calzado deportivo NOTA El conector de la manguera debe quedar a la altura del tobillo NOTA Consulte la imagen del manguito de p...