29
Preparation for Storage
If the unit is going to be stored in an off condition, allow the unit to warm up and dry out with the door open
before moving into storage.
Replacing Sensor (Chromatography Refrigerators)
The sensor needs replacement if the temperature goes below freezing in the cabinet. Use system alarms
to ensure the temperature inside the unit is always above 0°C. The sensors must be replaced if exposed to
freezing conditions.
If the display shows an error and sounds an alarm (in case of a control, defrost, bottle or ambient probe
failure), the sensor needs to be replaced.
Sensor Maintenance (Chromatography Refrigerators)
The sensor should be replaced every five years by a trained service provider.