NOTE:
If any of the following conditions are observed, the affected
module must be removed from use and inspected by qualified
personnel:
•
LED
segments are not illuminated on displays during
power-on test.
• Indicator lights do not illuminate.
• Appropriate module identification (
A
,
B
,
C
or
D
) is not
displayed.
If the affected module operates normally when it is attached
via the alternate
IUI
connector, it may be used until a
replacement module can be substituted.
1-4
General Setup and Operation
Alaris
®
System Directions for Use
PC Unit Section
Attaching and Detaching Modules
(Continued)
Detaching Module(s)
1.
Ensure module(s) is powered off before detaching.
2.
Push module release latch and then rotate module(s) up
and away from PC Unit or attached module (opposite to
motion shown above) to disengage connectors.
• Alaris
®
System reidentifies and shows appropriate
module identification (
A
,
B
,
C
or
D
), from left to right.
• Appropriate module position(s) (
A
,
B
or
C
) for
remaining module(s) appear on Main Display.
Add module as described in "Attaching Module(s)".
• System tests module, causing all
LED
segments and
indicator lights of displays to illuminate briefly.
• Appropriate module identification display (
A
,
B
,
C
or
D
)
illuminates. Modules are always labeled left to right, so if a
module is added to left of other modules, all modules will
be reidentified. Module reidentification does
NOT
interrupt
or affect infusion or monitoring on active modules.
• Module positions (
A
,
B
,
C
or
D
) appear on Main Display.
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......