2-66
Programming
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
5.
To begin multidose infusion, press
START
soft key.
• Main Display shows remaining
VTBI
for that dose.
• At completion of a multidose program,
MULTIDOSE
COMPLETE
appears on Main Display.
6.
To see detail screen during or between infusions, press
CHANNEL SELECT
key.
• During infusion,
Volume Remaining
displays.
• Between infusions:
♦
Number of doses completed and when next dose
starts displays.
♦
Yellow Standby Status Indicator illuminates.
Selecting Pressure Limit
1.
Press
CHANNEL SELECT
key.
2.
Press
OPTIONS
key.
SILENCE
OPTIONS
1
4
7
CLEAR
2
5
8
0
3
6
9
.
ENTER
CANCEL
SYSTEM
ON
Pump Module
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
NOTES:
Reference"Delay Options" procedure to program an infusion
delay. When delaying an infusion, a multidose cannot be
delayed for more than
8
hours, and all doses in the multidose
program must be completed within a
24
-hour program.
Syringe Module: If
NEOI
is enabled, the Near End of infusion
message appears near the end of the last dose.
Multidose Mode
(Continued)
Volume/Duration Disabled
(Continued)
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......